FDA Adverse Event Malfunction Summary report: N

2.5MM DRILL BIT/QC/GOLD/180MM

MDR report key: 17059452 · Received June 5, 2023

Report

Report Number
8030965-2023-07069
Event Type
Malfunction
Date Received
June 5, 2023
Date of Event
May 19, 2023
Manufacturer
SYNTHES GMBH
Product Code
GFA
UDI-DI
07611819018617
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D2B: ADDITIONAL DEVICE PRODUCT CODES: HSZ, GFF, HWE. D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. E3: REPORTER IS A J&J EMPLOYEE. H3, H4, H6: DEVICE HISTORY LOT. PART: 310.230. LOT: 41P6457. MANUFACTURING SITE: BETTLACH. RELEASE TO WAREHOUSE DATE: (B)(6) 2020. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 310.230, LOT: 41P6457, AND NO NON-CONFORMANCES WERE IDENTIFIED. INVESTIGATION SUMMARY: THE PHOTO WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED PHOTO. VISUAL ANALYSIS OF THE PHOTO REVEALED THAT THE DRILL BIT Ø2.5 L180/155 2FLUTE F/QUICK C WAS BROKEN FROM THE FLUTED TIP. THE BROKEN FRAGMENT WAS NOT VISIBLE IN THE PHOTO. NO OTHER ISSUES WERE FOUND. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED FOR DRILL BIT Ø2.5 L180/155 2FLUTE F/QUICK C. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED THAT ON MAY 19, 2023, THE 2.5 MM DRILL BIT FRACTURED DURING SURGERY. THERE WAS NO PATIENT HARM. THIS REPORT INVOLVES ONE 2.5MM DRILL BIT/QC/GOLD/180MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2314501 2.5MM DRILL BIT/QC/GOLD/180MM BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL GFA SYNTHES GMBH 41P6457 07611819018617

Patients

Seq Age Sex Outcome Treatment
1 Unknown