FDA Adverse Event Malfunction Summary report: N

CP5 DRIVE UNIT

MDR report key: 17058512 · Received June 5, 2023

Report

Report Number
9611109-2023-00252
Event Type
Malfunction
Date Received
June 5, 2023
Date of Event
May 8, 2023
Report Date
August 25, 2023
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWA
UDI-DI
04033817900948
PMA / PMN Number
K112225
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

FIELD SERVICE TECHNICIAN DISPATCHED TO THE FACILITY CONFIRMED THE ISSUE AND SOLVED IT BY REPLACING THE CP5 DRIVE UNIT. ALL FUNCTIONAL TESTS POSITIVELY PASSED AND THE UNIT WAS PUT BACK INTO SERVICE. COMPLAINTS DATABASE ANALYSIS REVEALED NO SIMILAR EVENT SINCE UNIT INSTALLATION IN 2021. BASED ON THE ABOVE, THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT IS ONE THE FOLLOWING: DEFECTIVE CPU (HKR 0325) WHICH IS LOCATED IN CP5 PUMP CONTROL PANEL; DEFECTIVE CIRCUIT BOARD MOTOR CONTROL (ZPR 0801) WHICH IS LOCATED IN THE DRIVE UNIT; COMMUNICATION FAILURE BETWEEN CPU (HKR 0325) AND MOTOR CONTROL BOARD (ZPR 0801). THE RISK IS IN THE ACCEPTABLE REGION. NO SPECIFIC ACTION WAS CURRENTLY DEEMED NECESSARY. LIVANOVA WILL KEEP MONITORING THE MARKET.

Additional Manufacturer Narrative · 0

A.1.-A.5. PATIENT INFORMATION WAS NOT PROVIDED. H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE SORIN CENTRIFUGAL PUMP 5 (CP5). THE INCIDENT OCCURRED IN UNITED STATES. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

SEE INTIAL REPORT.

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING A PROCEDURE, DESPITE THE RPM IN THE CP5 UNIT WERE FIXED, THEY OSCILLATED BETWEEN 500-600 RPM. THERE IS NO REPORT OF ANY PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1449209 CP5 DRIVE UNIT CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS DWA LIVANOVA DEUTSCHLAND 60-01-04 04033817900948

Patients

Seq Age Sex Outcome Treatment
1 Unknown