ATHLETIS PTA BALLOON DILATATION CATHETER
Report
- Report Number
- 2124215-2023-25787
- Event Type
- Malfunction
- Date Received
- June 2, 2023
- Date of Event
- May 15, 2023
- Report Date
- August 16, 2023
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- LIT
- UDI-DI
- 08714729974215
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
PRO CODE (PRODUCT CODE): DQY. DEVICE EVALUATED BY MFR.: ATHLETIS DEVICE WAS RECEIVED FOR ANALYSIS. A VISUAL EXAMINATION IDENTIFIED THAT THE BALLOON WAS NOT FOLDED WHICH INDICATES THAT THE DEVICE WAS SUBJECTED TO POSITIVE PRESSURE. A BALLOON PINHOLE WAS IDENTIFIED LOCATED AT THE DISTAL MARKER BAND. AS PER ATHLETIS SPECIFICATION THE RATED BURST PRESSURE FOR THIS DEVICE IS 30 ATMOSPHERES. A VISUAL EXAMINATION OBSERVED NO DAMAGE TO THE TIP OF THE DEVICE. A VISUAL AND TACTILE EXAMINATION FOUND NO KINKS OR DAMAGE TO THE SHAFT OF THE DEVICE. BOTH MARKER BANDS WERE UNDAMAGED AND PRESENT ON THE SHAFT OF THE DEVICE.
PRO CODE (PRODUCT CODE): DQY.
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN RADIAL ARTERY. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 25 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN RADIAL ARTERY. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 25 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1089232 | ATHLETIS PTA BALLOON DILATATION CATHETER | CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL | LIT | BOSTON SCIENTIFIC CORPORATION | 39347-100470 | 0030457742 | 08714729974215 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Female |