FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 17041014 · Received June 1, 2023

Report

Report Number
3008352382-2023-00092
Event Type
Malfunction
Date Received
June 1, 2023
Date of Event
May 2, 2023
Report Date
July 18, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420222
PMA / PMN Number
K141810
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: CATALOG: 442022; BATCH NO.: 3010404 & 2276304. CUSTOMER REPORTED A POSITIVE ID RESULT FOR BACTEC MEDIA, WHILE USING BIOFIRE FILMARRAY® BLOOD CULTURE IDENTIFICATION. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

REPORT 3 OF 4 IT WAS REPORTED THAT BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) MOLECULAR FALSE POSITIVE OF C. TROPICALIS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CULTURE RESULTS MATCH GRAM. BIOFIRE RESULTS GAVE EXTRA ¿FALSE POSITIVE¿- CANDIDA WE ARE HAVING URGENT MEETINGS REGARDING THE BD BOTTLES CAUSING A POSITIVE FLAG OF CANDIDA TROPICALIS WHEN WE USE THE BIOMERIUEX BCID PANELS. (B)(6) 2023 - ANAEROBIC - LOT 3010404 - BCID2 - RFIT-ASY-0147 - 2KL023 PATIENT (B)(6).

Description of Event or Problem · 0

REPORT 3 OF 4 IT WAS REPORTED THAT BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) MOLECULAR FALSE POSITIVE OF C. TROPICALIS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CULTURE RESULTS MATCH GRAM. BIOFIRE RESULTS GAVE EXTRA ¿FALSE POSITIVE¿- CANDIDA WE ARE HAVING URGENT MEETINGS REGARDING THE BD BOTTLES CAUSING A POSITIVE FLAG OF CANDIDA TROPICALIS WHEN WE USE THE BIOMERIUEX BCID PANELS. 4/26/2023 - ANAEROBIC - LOT 3010404 - BCID2 - RFIT-ASY-0147 - 2KL023 (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2323341 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 442022 3010404 00382904420222

Patients

Seq Age Sex Outcome Treatment
1 Unknown