FDA Adverse Event Injury Summary report: N

INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE

MDR report key: 17040013 · Received June 1, 2023

Report

Report Number
1037905-2023-00267
Event Type
Injury
Date Received
June 1, 2023
Date of Event
May 10, 2023
Report Date
June 1, 2023
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
PKL
UDI-DI
00827002580107
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K): K212323; INVESTIGATION EVALUATION: THE PRODUCT SAID TO BE INVOLVED WAS RETURNED IN A BIOHAZARD BAG WITH AN OPEN POUCH FROM THE LOT NUMBER PROVIDED IN THE REPORT. THE LABEL MATCHES THE PRODUCT RETURNED. OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE REPORT BASED ON THE CONDITION OF THE RETURNED DEVICE. A VISUAL EXAMINATION CONFIRMED THE CLIP HOUSING HAS SEPARATED FROM THE COIL CATHETER BUT REMAINS ATTACHED TO THE END OF THE DRIVE WIRE, IN A CLOSED POSITION. A FUNCTION TEST WAS ATTEMPTED, IN ORDER TO DEPLOY THE CLIP. WITH HANDLE MANIPULATION THE CLIP DEPLOYED, WITH SIGNIFICANT RESISTANCE ENCOUNTERED DURING DEPLOYMENT. THE DEVICE WAS THEN ADVANCED INTO THE ACCESSORY CHANNEL OF A PENTAX COLONOSCOPE (2.8MM CHANNEL) WHICH WAS PLACED IN A SIMULATED LOWER GI POSITION. THE TIP OF THE SCOPE WAS RETROFLEXED TO SIMULATE WORST CASE SCENARIO. WITH HANDLE MANIPULATION, THE DRIVE WIRE MOVED FREELY INSIDE THE OUTER SHEATH. A VISUAL EXAMINATION OF THE CLIP, CATHETER ATTACHMENT, AND COIL CATHETER (DISTAL END DEVICE COMPONENTS) SHOWED NO DEFORMITIES OR SIGNS OF DAMAGE. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF THE CLIP HOUSING DETACHING FROM THE CATH ATTACH FOR INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTIONS. THE INSTRUCTIONS FOR USE STATES, "NOTE: IF SEPARATION OF CLIP IS NOT IMMEDIATE, GENTLY MOVE CATHETER BACK AND FORTH OR USE OTHER ENDOSCOPIC MANEUVERS TO SEPARATE CATHETER FROM CLIP." PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

DURING A PROCEDURE TO CLIP A BLEEDING POST POLYPECTOMY, THE PHYSICIAN USED A COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE. IT WAS REPORTED THAT THE CLIP CLOSED ON THE TISSUE BUT WAS UNABLE TO DEPLOY OR REOPEN SO THE PHYSICIAN HAD TO PULL THE CLIP AND CATHETER AWAY FROM THE SITE CAUSING MORE BLEEDING. AFTER THE PULLED THEY UNDEPLOYED CLIP FROM THE TISSUE THEY PLACED ANOTHER CLIP TO STOP THE INCREASED BLEEDING SITE. NO ADVERSE EFFECT ON THE PATIENT WAS REPORTED DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1029450 INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE PKL, LIGATOR, HEMORRHOIDAL PKL WILSON-COOK MEDICAL INC G58010 W4634983 00827002580107

Patients

Seq Age Sex Outcome Treatment
1 Unknown SCOPE, UNKNOWN MAKE AND MODEL