FDA Adverse Event
Malfunction
Summary report: N
ENZTEC CLASSIC OCI - CORIN TRINKET
MDR report key: 17037625
·
Received June 1, 2023
Report
- Report Number
- 3004641308-2023-00001
- Event Type
- Malfunction
- Date Received
- June 1, 2023
- Date of Event
- April 24, 2023
- Report Date
- June 1, 2023
- Manufacturer
- ENZTEC LIMITED
- Product Code
- HWA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 0
DURING CUP IMPACTION OF A THR SURGERY IN AUSTRALIA, THE OFFSET CUP IMPACTOR TRINKET THREAD BECAME STUCK IN THE DEFINITIVE CUP IMPLANT. SURGEON ATTEMPTED TO UNSCREW THE IMPACTOR USING THE SCREWDRIVER ATTACHMENT AND THE CUP DISLODGED. SURGEON REAMED UP A SIZE AND IMPACTED CUP IMPLANT USING AN ALTERNATE BACKUP INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1830492 | ENZTEC CLASSIC OCI - CORIN TRINKET | OCI TRINKET | HWA | ENZTEC LIMITED | 3991055001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Female | Required Intervention | ENZTEC CLASSIC OCI - CORIN |