FDA Adverse Event Malfunction Summary report: N

BD GLIDE¿ WITH TBL BD ULTRA-FINE¿ NEEDLE INSULIN SYRINGE

MDR report key: 17035630 · Received May 31, 2023

Report

Report Number
1920898-2023-00324
Event Type
Malfunction
Date Received
May 31, 2023
Date of Event
May 8, 2023
Report Date
August 31, 2023
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE VIDEO PROVIDED. THE CUSTOMER RETURNED A VIDEO OF THE 31GX6MM BD SYRINGE. THE CUSTOMER REPORTED THAT WHEN THE INSULIN SYRINGE CAP WAS OPENED THERE WAS A YELLOWISH PART NEAR NEEDLE END. THE VIDEO WAS EXAMINED, AND THERE APPEARS TO BE A BROWNISH COLOR STAIN AROUND THE RIM/TIP OF THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 2052405. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. BASED ON THE PHOTOS RECEIVED, EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE (FOREIGN MATTER ON BARREL TIP). THERE WERE NO QUALITY NOTIFICATIONS OR DISPATCHES THAT PERTAINED TO THE COMPLAINT; THEREFORE, NO ROOT CAUSE CAN BE DETERMINED.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 300 BD GLIDE¿ WITH TBL BD ULTRA-FINE¿ NEEDLE INSULIN SYRINGES' CAPS WERE DISCOLORED TO YELLOW NEAR THE NEEDLE POINT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN THEY OPEN INSULINE SYRINGE CAP YELLOWISH PART NEAR NEEDLE END."

Description of Event or Problem · 0

IT WAS REPORTED THAT 300 BD GLIDE¿ WITH TBL BD ULTRA-FINE¿ NEEDLE INSULIN SYRINGES' CAPS WERE DISCOLORED TO YELLOW NEAR THE NEEDLE POINT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN THEY OPEN INSULINE SYRINGE CAP YELLOWISH PART NEAR NEEDLE END."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1682141 BD GLIDE¿ WITH TBL BD ULTRA-FINE¿ NEEDLE INSULIN SYRINGE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 2052405

Patients

Seq Age Sex Outcome Treatment
1 Unknown