FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM

MDR report key: 17034881 · Received May 31, 2023

Report

Report Number
0002023141-2023-01516
Event Type
Injury
Date Received
May 31, 2023
Date of Event
December 28, 2022
Report Date
October 30, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019508
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED IMPLANT (TSVB10) WAS NOT RETURNED FOR INVESTIGATION. THEREFORE, VISUAL EVALUATION COULD NOT BE PERFORMED. THE INVESTIGATION WAS CONDUCTED USING APPLICABLE INSTRUCTIONS FOR USE, RISK FILES AND OTHER AVAILABLE INFORMATION. DHR REVIEW FOR THE LOT (1244048) HAD REVEALED NO DEVIATIONS NOR NON-CONFORMANCES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. ALL PRODUCTS WERE CONFORMING AT THE TIME THEY LEFT ZIMVIE. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT (1244048) AND 1 SIMILAR EVENT OR COMPLAINT WAS FOUND (B)(4). (COMPLAINT CATEGORY KEYWORD: FRACTURE: IMPLANT). THE CUSTOMER DID NOT SUBMIT ANY IMAGES/X-RAY. REVIEW OF APPROPRIATE DOCUMENTATION: DOCUMENTS REVIEWED: INSTRUCTIONS FOR USE - TAPERED SCREW-VENT AND TRABECULAR METAL IMPLANTS, IFU4869 REV 9-10/19. INFORMATION IDENTIFIED: CONTRAINDICATIONS, PRECAUTIONS, AND ADVERSE EFFECTS. PER THE APPLICABLE IFU, IT IS STATED THAT IMPROPER TECHNIQUES, SEVERE BRUXISM, CLENCHING, AND OVERLOADING, MAY CAUSE COMPONENT FRACTURE. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS LONG-TERM PARAFUNCTIONAL HABITS (E.G., CLENCHING, BRUXISM, AND OVERLOADING) OF THE PATIENT OVER THE IMPLANTATION PERIOD. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION AND THE REPORTED EVENT COULD NOT BE VERIFIED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT G3: DATE RECEIVED BY MANUFACTURER G6: CHECKED "FOLLOW-UP" H2: CHECKED FOLLOW-UP TYPE H3: CHANGED "YES" TO "NO" H6: ENTERED EVALUATION CODES H10: ADDED MANUFACTURER NARRATIVE

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED G4: K011028, K013227.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANTS AT TOOTH SITE #19 WAS REMOVED DUE TO FRACTURE. PATIENT WAS HAVING PAIN IN IMPLANT SITE BECAUSE SHE WAS CHEWING SOME FOOD AND IT FRACTURED IMPLANTS. REQUESTED TO BE REMOVED. SYMPTOMS AS A RESULT OF THE EVENT: PAIN

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1557704 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB10 1244048 00889024019508

Patients

Seq Age Sex Outcome Treatment
1 61 YR Prefer Not To Disclose Other