START-X TIP EMS INSERT 2
Report
- Report Number
- 8031010-2023-00795
- Event Type
- Malfunction
- Date Received
- May 31, 2023
- Date of Event
- May 17, 2023
- Report Date
- July 24, 2023
- Manufacturer
- MAILLEFER INSTRUMENTS HOLDING SARL
- Product Code
- ELC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
SUMMARY: RETURNED START-X TIP EMS INSERT 2 IS ACTUALLY BROKEN IN THE ACTIVE PART. NO MATERIAL DEFECT WAS FOUND DURING ANALYSIS OF THE RUPTURE PATTERN. NO UNUSED DEVICE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1772251). NO INFORMATION WAS GIVEN REGARDING TECHNIQUE, WE CANNOT RULE ON ITS COMPLIANCE WITH MAILLEFER'S RECOMMENDATIONS. ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND.
WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, IF THIS MALFUNCTION RECURRED, IT COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
IN THIS EVENT IT IS REPORTED THAT START-X TIP EMS INSERT 2 BROKE DURING USE. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1836877 | START-X TIP EMS INSERT 2 | SCALER, ULTRASONIC | ELC | MAILLEFER INSTRUMENTS HOLDING SARL | 1772251 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |