FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES

MDR report key: 17028472 · Received May 30, 2023

Report

Report Number
9617032-2023-00626
Event Type
Malfunction
Date Received
May 30, 2023
Date of Event
May 3, 2023
Report Date
May 19, 2023
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903679558
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B3: DATE OF EVENT IS UNKNOWN. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.6 INVESTIGATION SUMMARY. MATERIAL #: 367955. LOT/BATCH #: 2123429. BD HAD NOT RECEIVED SAMPLES, BUT ONE (1) PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR ADDITIVE ABNORMALITY WITH THE INCIDENT LOT WAS NOT OBSERVED. EVALUATION OF THE PHOTOGRAPH INDICATED A SHELF PACK OF UNUSED TUBES WITH NORMAL SILICA SPRAY ON THE INNER TUBES WALLS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE ADDITIVE ABNORMALITY. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES THAT THERE WAS DISCOLORED AND ABNORMAL ADDITIVE FORM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE MATERIAL DELIVERY HAS BEEN DECLINED BECAUSE THE BATCH PRESENTS OBSERVATIONS IN PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1029593 BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 367955 2123429 50382903679558

Patients

Seq Age Sex Outcome Treatment
1 Unknown