INTELLIS
Report
- Report Number
- 3004209178-2023-07652
- Event Type
- Malfunction
- Date Received
- May 30, 2023
- Date of Event
- October 31, 2022
- Report Date
- May 30, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LGW
- UDI-DI
- 00643169781702
- PMA / PMN Number
- P840001
- Removal / Correction Number
- Z-1535-2021
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
CONTINUATION OF D10: PRODUCT ID 97755 LOT# SERIAL# (B)(6) PRODUCT TYPE RECHARGER SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 97755, SERIAL/LOT #: (B)(6), UDI#: (B)(6) H3: ANALYSIS OF THE 97755 RECHARGER (RTM) (S/N (B)(6)) REVEALED NON-CONFORMANCES. THE RECHARGE ANTENNA FAILED DUE A "NO DEVICE FOUND" ERROR MESSAGE. THE RECHARGE ANTENNA FAILED DUE TO THE RMS VOLTAGE AT THE H FIELD PROBE. THIS REGULATORY REPORT IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW AS PART OF A CAPA 620188 ACTION. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT WHEN THE PATIENT (PT) WAS TRYING TO CONNECT THE CONTROLLER TO THE IMPLANTABLE NEUROSTIMULATOR (INS), IT WAS NOT CONNECTING TO CHARGE. CALLER SAID IT SEEMED LIKE THERE WAS A LOOSE WIRE BECAUSE THE NUMBERS WERE JUMPING AROUND WHEN TRYING TO CHARGE. CALLER THEN HAD PT TRY CHARGING THEIR INS ON THE CALL AND PT WAS SEEING NO DEVICE FOUND. PT ENTERED A PASSIVE RECHARGING MODE (PRM) AND HAD NUMBERS STARTED ALL AT 0 AND THEN THE MIDDLE NUMBER WAS JUMPING ANYWHERE FROM 18 TO 100. THE RECHARGER CORD THEN GOT VERY HOT. A REPLACEMENT RECHARGER (RTM) WAS SENT OUT. NO SYMPTOMS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1159600 | INTELLIS | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR | LGW | MEDTRONIC PUERTO RICO OPERATIONS CO. | 97715 | 00643169781702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | Female |