FDA Adverse Event Injury Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 17027674 · Received May 30, 2023

Report

Report Number
3004753838-2023-106886
Event Type
Injury
Date Received
May 30, 2023
Date of Event
April 5, 2023
Report Date
May 30, 2023
Manufacturer
DEXCOM, INC.
Product Code
QBJ
UDI-DI
80386270000251
PMA / PMN Number
DEN170088
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). DIABETES MELLITUS IS A KNOWN CAUSE OF DIABETIC KETOACIDOSIS. HEALTH EFFECT - CLINICAL CODE - MYOCARDIAL INFARCTION - 1969 - E061202. HEALTH EFFECT - CLINICAL CODE - ISCHEMIC HEART DISEASE - E0612.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. ON (B)(6) 2023, THE PATIENT¿S CGM DISPLAYED LOW AND HE SELF-TREATED WITH FOOD AND JUICE; HOWEVER, THE CGM CONTINUED TO DISPLAY LOW. THE PATIENT BECAME ¿OUT OF IT¿ AND HIS SPOUSE CALLED EMERGENCY MEDICAL SERVICES. EMS TRANSPORTED THE PATIENT TO THE HOSPITAL WHERE A BG WAS PERFORMED WHICH WAS 1300 MG/DL. THE PATIENT WAS DIAGNOSED WITH DIABETIC KETOACIDOSIS (DKA) AND A MYOCARDIAL INFARCTION (MI) WHICH WAS ASSOCIATED WITH HIS BG OF 1300 MG/DL. HE WAS ADMITTED TO THE INTENSIVE CARE UNIT (ICU), TREATED WITH IV INSULIN (NOVOLOG) AND FIVE OTHER UNSPECIFIED IV MEDICATIONS. AFTER 5 DAYS IN THE ICU, HIS BG RETURNED TO NORMAL AND HE WAS TRANSFERRED TO A STEP-DOWN UNIT. AFTER A TOTAL OF 7 DAYS, THE PATIENT WAS DISCHARGED FROM THE HOSPITAL. AT THE TIME OF THE REPORT, THE PATIENT RECOVERED, WAS HOME AND REPORTED THAT THEY GET TIRED EASILY. DATA WAS EVALUATED AND THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1829870 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. STS-OM-001 NI 80386270000251

Patients

Seq Age Sex Outcome Treatment
1 74 YR Male Other| H