FDA Adverse Event Injury Summary report: N

BL, Ø 4.1MM RC, SLA® 12MM, TI, LOXIM

MDR report key: 17026453 · Received May 30, 2023

Report

Report Number
0009613348-2023-055087
Event Type
Injury
Date Received
May 30, 2023
Date of Event
November 4, 2022
Report Date
May 30, 2023
Manufacturer
INSTITUT STRAUMANN AG
Product Code
DZE
UDI-DI
07630031707000
PMA / PMN Number
K062129
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BATCH NUMBER COULD BE VERIFIED. OUR MANUFACTURING Q-SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFIRM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED.  THE NON-INTEGRATION OF AN ENDOSSEOUS DENTAL IMPLANT DURING THE HEALING PHASE IS A KNOWN INHERENT RISK OF THE TREATMENT WITH DENTAL IMPLANTS. MOST IMPLANT FAILURES OCCUR BEFORE OCCLUSAL LOADING (ANALAINEN ET AL. 2009). BASED ON CLINICAL STUDIES ABOUT 1-3% OF THE IMPLANTS FAIL WITHIN THE FIRST YEAR AFTER IMPLANTATION (GANELES ET AL. 2008). IMPLANTS MAY HAVE TO BE REMOVED IN CASE ONE OR MORE OF THE IMPLANT SUCCESS CRITERIA ARE NOT MET. IMPLANT SUCCESS CRITERIA ACCORDING TO BUSER ET AL. (1991) ARE: 1. ABSENCE OF PERSISTENT SUBJECTIVE COMPLAINTS SUCH AS PAIN, FOREIGN, BODY SENSATION AND /OR DYSESTHESIA, 2. ABSENCE OF A RECURRENT PERI-IMPLANT INFECTION WITH SUPPURATION, 3. ABSENCE OF IMPLANT MOBILITY, 4. ABSENCE OF A CONTINUOUS RADIOLUCENCY AROUND THE IMPLANT. THE MANUFACTURER'S TREND ANALYSIS CONFIRMS THAT THE REPORTED EARLY FAILURE RATE ASSOCIATED WITH ITS DENTAL IMPLANTS IS BELOW THE EXPECTED FAILURE RATE FOR THIS TREATMENT AS PUBLISHED IN THE SCIENTIFIC LITERATURE.

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED 2022-11-03 IN FDI 26. ON 2022-11-04, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1849160 BL, Ø 4.1MM RC, SLA® 12MM, TI, LOXIM ENDOSSEOUS DENTAL IMPLANT DZE INSTITUT STRAUMANN AG SLA Titanium BL GPW76 07630031707000

Patients

Seq Age Sex Outcome Treatment
1 45 YR Female Required Intervention