IQECG
Report
- Report Number
- 2081230-2010-00001
- Event Type
- Malfunction
- Date Received
- May 17, 2010
- Date of Event
- March 30, 2010
- Report Date
- May 14, 2010
- Manufacturer
- BRENTWOOD MED TECHNOLOGY DBA MIDMARK DIAGNOSTICS GROUP
- Product Code
- DSI
- PMA / PMN Number
- K955023
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE IQECG SOFTWARE IS DESIGNED TO CALCULATE THE QTC VALUE UTILIZING BAZETT'S EQUATION FOR PATIENTS (B)(6), AND HODGES EQUATION FOR PATIENTS (B)(6). (B)(6) DETERMINED THAT WHEN THE USER ACCESSES THE QT DETAILS OPTION TO EITHER EDIT OR ACCEPT THE QT INTERVAL, THE IQECG SOFTWARE RECALCULATES THE QTC VALUE USING THE HODGES EQUATION, REGARDLESS OF PATIENT'S AGE, WITHOUT NOTIFYING THE USER. THE HODGES EQUATION RESULTS A LOWER QTC VALUE THAN THE BAZETT'S EQUATION. THE RECALCULATION ONLY OCCURS WHEN THE USER (LIKELY SPECIALIZED PEDIATRIC CARDIOLOGIST) MANUALLY EDITS OR ACCEPTS THE QT. PUBLISHED ARTICLES ON THE QTC SUBJECT SHOW THAT BOTH HODGES AND BAZETT'S EQUATIONS ARE WIDELY USED AND THERE IS DISAGREEMENT ON WHICH EQUATION IS BETTER. THE ARTICLES DISCUSSING THE USE OF QTC IN LONG QT SYNDROME (LQTS) DIAGNOSIS EMPHASIZE THAT QTC MEASUREMENTS SHOULD BE INTERPRETED IN THE CONTEXT OF THE PATIENT'S PERSONAL AND FAMILY HISTORY (BJSM, 2009 43: 657-662) AND THAT MANY PATIENTS COULD BE DIAGNOSED INACCURATELY WITH LQTS IF THE DIAGNOSIS WAS MADE PRINCIPALLY BECAUSE OF A BORDERLINE QTC MEASUREMENT (ANN PEDIATR CARD 2008 VOL 1 ISSUE 1). THESE STATEMENTS ARE CONSISTENT WITH IQECG WARNING STATEMENT, "NOT ALL ECG ABNORMALITIES CAN BE DETECTED BY COMPUTERIZED / AUTOMATED ECG ANALYSIS ALGORITHM. THE SUGGESTED INTERPRETATION, INCLUDING NUMERICAL AND GRAPHICAL RESULTS, SHOULD BE EXAMINED WITH RESPECT TO THE PATIENTS OVERALL CLINICAL CONDITION." (B)(6) CONSULTED WITH EXPERIENCED PEDIATRIC CARDIOLOGISTS WHO OFFER OPINION THAT THE PROBABILITY FOR HARM IS VERY LOW.
USING IQECG SOFTWARE VERSION 8.0.1, THE USER DOUBLE CLICKED ON THE QT VALUE ON THE ECG EDIT SCREEN TO EDIT THE QT INTERVAL. WHEN THE USER DID NOT CHANGE THE QT INTERVAL AND CLICKED "OK", THE QTC VALUE RECALCULATED. THE QTC VALUE CHANGED FROM 423 MSEC TO 397 MSEC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IQECG | DSI | BRENTWOOD MED TECHNOLOGY DBA MIDMARK DIAGNOSTICS GROUP | IQECG | NONE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |