ELECSYS 2010 RACK
Report
- Report Number
- 1823260-2010-03230
- Event Type
- Malfunction
- Date Received
- May 28, 2010
- Date of Event
- May 11, 2010
- Report Date
- August 10, 2010
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K961481
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GB
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.
A DEFINITIVE ROOT CAUSE FOR THE ISSUE COULD NOT BE DETERMINED. THE MOST LIKELY CAUSE WAS A SAMPLE PIPETTING FAILURE, POSSIBLY CAUSED BY THE PROBE AND MIXER PADDLE ISSUES FOUND. ANOTHER CAUSE OF A PIPETTING ISSUE COULD ALSO BE NON-VERTICAL PLACEMENT OF THE SAMPLE TUBE IN THE SAMPLING RACK.
THE CUSTOMER ALLEGED QUESTIONABLE BETA-HCG RESULTS WERE OBTAINED ON 2 PATIENTS: PATIENT A: SAMPLE FROM (B) (6) 2010: HCG = 14,225 RERUN, AFTER RESPINNING SAMPLE: HCG = 221 ************************ PATIENT B: SAMPLE FROM (B) (6) 2010 AT 10:00: HCG = 44 (RESULT RELEASED) THIS SAMPLE WAS NOTED AS "MERKY" BY THE CUSTOMER. RERUN, AFTER RESPINNING SAMPLE: HCG = 17,877 ************************** HCG-STAT REAGENT LOT #: 156661 NO ADVERSE EVENT HAS BEEN ALLEGED REGARDING THESE DISCREPANCIES. AN ENGINEER VISIT FOUND PROBLEMS WITH THE S/R PROBE AND MIXER PADDLE, WHICH WERE REPLACED. THE CUSTOMER BELIEVES THAT THE PROBLEM HAS BEEN RESOLVED.
THERE WERE FOUR ENDEAVOR SPRINT OVER THE WIRE(OTW) DRUG-ELUTING STENTS IMPLANTED, ONE OF THE LEFT MAIN, ONE IN THE MID RCA AND TWO IN THE RIGHT PDA (REFERENCE MFR #'S 2853200-2010-01501, 01502, 01503). AT 30 DAY FOLLOW UP, PATIENT'S CARDIAC STATUS WAS STABLE ANGINA. PATIENT WAS ASYMPTOMATIC / FREE OF SYMPTOMS AT 6 MONTH FOLLOW UP. APPROXIMATELY ONE YEAR POST INDEX PROCEDURE, PATIENT WAS WORKING AND GOT REAL SWEATY AND HAD SHORTNESS OF BREATH AND DEVELOPED CHEST HEAVINESS RADIATING TO SHOULDERS. IT WAS REPORTED THAT THE PATIENT PRESENTED TO LOCAL EMERGENCY ROOM. PATIENT WAS PRESCRIBED NOTIR AND MORPHINE. IT IS REPORTED THAT AN ACUTE NON-STEMI OCCURRED AND A REVASCULARIZATION OF TWO LESIONS WAS CARRIED OUT. AT 12 MONTH FOLLOW UP, PATIENT'S CARDIAC STATUS WAS UNSTABLE ANGINA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ELECSYS 2010 RACK | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |