FDA Adverse Event Malfunction Summary report: N

LIBRE 2 SENSOR FREESTYLE

MDR report key: 17016357 · Received May 29, 2023

Report

Report Number
2954323-2023-22130
Event Type
Malfunction
Date Received
May 29, 2023
Date of Event
May 18, 2023
Report Date
October 9, 2023
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
QLG
UDI-DI
00357599800000
PMA / PMN Number
K193371
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SENSOR (B)(6) HAS BEEN RETURNED AND INVESTIGATED. THE SENSOR PLUG WAS FULLY SEATED AND NO ISSUES WERE OBSERVED. EXTRACTED DATA FROM THE RETURNED SENSOR USING APPROVED SOFTWARE. THE SENSOR WAS FOUND TO BE IN SENSOR STATE 5 (INDICATING NORMAL TERMINATION). VISUAL INSPECTION WAS PERFORMED ON PCBA(PRINTED CIRCUIT BOARD ASSEMBLY) TRIANGLE AND OBSERVED DEBRIS FLAKES OF MATERIAL. THE DEBRIS OBSERVED WAS MOSTLY LIKELY CAUSED WHEN THE PLUG WAS REMOVED DURING PRODUCT RETURN INVESTIGATION. CURRENT WAS APPLIED TO THE SENSOR TO PERFORM ACCURACY TESTING WHILE IN THE TEST FIXTURE. ALL RESULTS WERE WITHIN SPECIFICATION. POISE VOLTAGE AND SENSOR THERMISTOR TESTING WERE BOTH WITHIN SPECIFICATION, INDICATING THE SENSOR WAS PROVIDING ACCURATE GLUCOSE READINGS. PQE PHYSICALLY INSPECTED THE RETURNED PUCK AND REVIEWED THE ATTACHMENT FROM THE PHASE 1 INVESTIGATION, AND OBSERVED DEBRIS FLAKES OF MATERIAL IN THE PLUG TRIANGLE AREA. PQE DETERMINED THE DEBRIS FLAKES OF MATERIAL OBSERVED WERE NOT SEVERE ENOUGH TO HAVE ANY IMPACT ON THE SENSORS ACCURACY. NO ISSUES WERE FOUND DURING TESTING, THEREFORE, THIS ISSUE IS NOT CONFIRMED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED. VISUAL INSPECTION WAS PERFORMED ON RETURNED SENSOR AND NO ISSUES WERE OBSERVED. PQE OBSERVED THE TEST TO PASS IN PHASE 1. THE RETURNED SENSOR WAS REPROGRAMMED AND PERFORMED ACCURACY TEST. ALL RESULTS WERE WITHIN SPECIFICATION. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. AT THIS TIME PRODUCT HAS NOT YET BEEN RETURNED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DHRS FOR THE FS LIBRE SENSOR AND FS LIBRE SENSOR KIT WERE REVIEWED AND THE DHRS SHOWED THE FS LIBRE SENSOR AND SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
850106 LIBRE 2 SENSOR FREESTYLE FLASH GLUCOSE MONITORING SYSTEM QLG ABBOTT DIABETES CARE INC 71992-01 00357599800000

Patients

Seq Age Sex Outcome Treatment
1 Female