FDA Adverse Event Malfunction Summary report: N

SURGERY TABLE

MDR report key: 17015 · Received October 20, 1994

Report

Report Number
MW1003746
Event Type
Malfunction
Date Received
October 20, 1994
Date of Event
September 19, 1994
Report Date
October 10, 1994
Manufacturer
AMERICAN STERILIZER CO.
Product Code
FQO
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

PT LYING SUPINE ON OR TABLE. ONE NURSE AT FOOT OF BED HOLDING PT'S LEGS AND ONE NURSE AT PT'S LEFT SIDE ATTACHING LEG HOLDER TO TABLE. TABLE TOP SUDDENLY TILTED TO RIGHT APPROX 45 DEGREES. PT WAS GRABBED BY NURSES, TABLE TOP LEVELED AND PT TRANSFERRED TO OR CART, THEN TO ANOTHER SURGERY TABLE. FURTHER INVESTIGATION SHOWED THAT WHEN THE MFR WAS INSTALLING AN UPGRADE KIT TO PREVENT THIS, A LOCKING SCREW WAS OMITTED. PROBLEM HAS BEEN CORRECTED, AND OTHER TABLES CHECKED FOR SIMILAR OMISSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SURGERY TABLE SURGERY TABLE FQO AMERICAN STERILIZER CO. 3080RL (BL58728200)

Patients

Seq Age Sex Outcome Treatment
1 * Other