FDA Adverse Event Malfunction Summary report: N

APPOSE ULC

MDR report key: 17013470 · Received May 26, 2023

Report

Report Number
9612501-2023-00878
Event Type
Malfunction
Date Received
May 26, 2023
Date of Event
April 24, 2023
Report Date
May 26, 2023
Manufacturer
DAVIS & GECK CARIBE LTD
Product Code
GDT
UDI-DI
10884521104365
PMA / PMN Number
K900486
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D10 CONCOMITANT PRODUCT: 8886803512 APPOSE ULC 35 REG SKIN STAP (LOT#: J2F1762LY); 8886803512 APPOSE ULC 35 REG SKIN STAP (LOT#: J2F1762LY); 8886803512 APPOSE ULC 35 REG SKIN STAP (LOT#: J2F1762LY); 8886803512 APPOSE ULC 35 REG SKIN STAP (LOT#: J2F1762LY) MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING STAPLING SKIN PROCEDURE, MULTIPLE BOXES OF STAPLERS WERE USED ON A PEDIATRIC BURN PATIENT TO STAPLE SKIN GRAFT, BUT AFTER THE FIRST FEW STAPLES WERE USED, 5 STAPLERS FROM THE SAME LOT WERE JAMMED. THE STAPLE PARTIALLY DEPLOYED, BUT THE STAPLE WAS STUCK IN THE END OF THE STAPLER. SOME WERE PULLED AND "UNJAM" THE STAPLER, BUT IT WAS NOT POSSIBLE FOR OTHER. THE SURGICAL TIME WAS EXTENDED FOR 30 MINUTES OR MORE. ANOTHER STAPLER WAS OPENED TO RESOLVE THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1728713 APPOSE ULC STAPLE, REMOVABLE (SKIN) GDT DAVIS & GECK CARIBE LTD 8886803512 J2F1762LY 10884521104365

Patients

Seq Age Sex Outcome Treatment
1 Unknown SEE NOTE ON H10