LVIS EVO
Report
- Report Number
- 2032493-2023-00749
- Event Type
- Malfunction
- Date Received
- May 26, 2023
- Date of Event
- May 15, 2023
- Report Date
- May 22, 2023
- Manufacturer
- MICROVENTION, INC.
- Product Code
- QCA
- UDI-DI
- 00842429103449
- PMA / PMN Number
- P170013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ITEMS RETURNED FOR INVESTIGATION: PUSHER, MICROCATHETER. ITEMS NOT RETURNED FOR INVESTIGATION: STENT, INTRODUCER. THE PUSHER WAS RETURNED BROKEN AT THE DISTAL TIP. THE MICROCATHETER WAS WAVY AT 50CM FROM THE DISTAL TIP. THE PUSHER DISTAL TIP WAS EXAMINED UNDER HIGH MAGNIFICATION CONFIRMING THAT THE TIP WAS BROKEN.
A SEARCH FOR NON-CONFORMANCES ASSOCIATED WITH THE REPORTED PART/LOT NUMBER COMBINATION DID NOT REVEAL ANY PRODUCTION-RELATED ISSUES RELEVANT TO THE COMPLAINT THAT OCCURRED DURING MANUFACTURING OF THE DEVICE. THE DEVICE IS REPORTED AVAILABLE BUT HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR ANALYSIS. THEREFORE, THE ALLEGED PRODUCT ISSUE CANNOT BE CONFIRMED. IF THE DEVICE OR ADDITIONAL INFORMATION IS RECEIVED, MICROVENTION, INC., WILL SUBMIT A SUPPLEMENTAL REPORT.
IT WAS REPORTED THAT DURING TREATMENT OF A PATIENT ENROLLED IN THE SELANT STUDY, AN LVIS EVO WAS USED FOR A STENT ASSISTED COILING CASE. IT WAS OBSERVED THAT FOLLOWING SUCCESSFUL DEPLOYMENT OF THE STENT AND SUBSEQUENT WITHDRAWAL OF THE DELIVERY WIRE, THE RADIO-OPAQUE TIP DETACHED FROM THE DELIVERY WIRE. A FOLLOW UP POST PROCEDURE X-RAY AND MR CONFIRMED THERE WERE NO CLINICAL ISSUES. ADDITIONAL INFORMATION INDICATED THE RADIO-OPAQUE TIP REMAINED WITHIN THE STENT. NO INTERVENTION WAS TAKEN TO REMOVE THE TIP. THE PATIENT WAS REPORTED TO BE WELL AND DISCHARGED FROM THE HOSPITAL. NO ADDITIONAL TREATMENT WAS PLANNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1728641 | LVIS EVO | INTRALUMINAL DEVICES | QCA | MICROVENTION, INC. | LEV2522-MVE | 0000327229 | 00842429103449 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |