FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 17011705 · Received May 26, 2023

Report

Report Number
1221359-2023-01159
Event Type
Malfunction
Date Received
May 26, 2023
Date of Event
May 16, 2023
Report Date
June 13, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 206540 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 206540 AND DEVICE PART NUMBER 195-430WJRH / LOT 202964. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE AND/OR FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 206540 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED COMPLETE. ABBOTT DIAGNOSTICS SCARBOROUGH WILL CONTINUE TO MONITOR AND TREND FOR THIS REPORTED ISSUE AS PART OF OUR ONGOING POST MARKET SURVEILLANCE. H3 OTHER TEXT : DEVICE DISCARDED, SINGLE USE

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : DEVICE DISCARDED, SINGLE USE.

Description of Event or Problem · 0

THE CONSUMER REPORTED A CONFLICTING RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON 16MAY2023. THE CONSUMER RECEIVED A POSITIVE AND A NEGATIVE RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON THE SAME DAY. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A CONFLICTING RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON (B)(6)2023. THE CONSUMER RECEIVED A POSITIVE AND A NEGATIVE RESULT WITH BINAXNOW COVID-19 AG SELF-TEST ON THE SAME DAY. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1437062 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 206540 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female