OLYMPUS HIGH FREQUENCY RESECTION ELECTRODE
Report
- Report Number
- 9610773-2010-00012
- Event Type
- Other
- Date Received
- May 24, 2010
- Date of Event
- April 23, 2010
- Report Date
- April 26, 2010
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- JOS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- NURSE
Narratives
THE SUBJECT DEVICE REFERENCED IN THIS REPORT WAS RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION CONFIRMED THE USERS REPORT, AS THE TIP OF THE ELECTRODE WAS FOUND DAMAGED AND DETACHED. THERE WAS EVIDENCE OF THERMAL DAMAGE NOTED ON BOTH ENDS OF THE LOOP PIECES, AND BIOMATERIALS ON THE SURFACE OF THE LOOP. THE EXACT CAUSE OF THE USER'S EXPERIENCE COULD NOT BE CONCLUSIVELY DETERMINED. NO OTHER EQUIPMENT USED DURING THE PROCEDURE WAS RETURNED FOR EVALUATION, AS THERE WERE NO ALLEGATIONS OF MALFUNCTION ASSOCIATED WITH THE OTHER ITEMS. THE DEVICE WILL BE FORWARDED TO THE ORIGINAL EQUIPMENT MANUFACTURER FOR FURTHER EVALUATION. IF SIGNIFICANT ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE PROVIDED. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
THE USER FACILITY REPORTED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF PROSTATE (TURP) THE DISTAL TIP OF THE ELECTRODE LOOP BROKE OFF INSIDE THE PT AFTER TWENTY MINUTES OF USE. THE BROKEN PORTION OF THE LOOP WAS RETRIEVED WITH AN UNK DEVICE. THE PROCEDURE WAS COMPLETED USING A DIFFERENT, BUT SIMILAR ELECTRODE. THERE WAS NO PT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS HIGH FREQUENCY RESECTION ELECTRODE | LOOP ELECTRODE | JOS | OLYMPUS WINTER & IBE GMBH | WA22302D | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | WORKING ELEMENT WA22367A| TELESCOPE MODEL A2201A| OUTER SHEATH A22026A| INNER SHEATH A22040A| HF CABLE WA00013A| OLYMPUS UES-40 SERIAL NUMBER (B)(4) |