FDA Adverse Event Injury Summary report: N

KIT IMPLANTABLE SLIM TIP LEAD, 50CM

MDR report key: 17010420 · Received May 26, 2023

Report

Report Number
1627487-2023-02399
Event Type
Injury
Date Received
May 26, 2023
Date of Event
April 13, 2023
Report Date
July 6, 2023
Manufacturer
ABBOTT MEDICAL
Product Code
PMP
UDI-DI
05415067027153
PMA / PMN Number
P150004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.

Additional Manufacturer Narrative · 0

DATE OF EVENT IS ESTIMATED. THE ALLEGATION IS AGAINST 1 OF 2 LEADS; HOWEVER, IT IS UNKNOWN WHICH LEAD, THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: 50CM IMPLANT LEAD KIT, SLIM TIP, MODEL: MN10450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 6840452.

Additional Manufacturer Narrative · 0

CORRECTION: SECTION D4, D6A, AND H4 HAVE BEEN UPDATED TO REFLECT THE CORRECT DEVICE INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WILL BE UNDERTAKEN IN THE FUTURE TO ADDRESS THIS ISSUE. IT IS UNKNOWN WHICH LEAD HAS HIGH IMPEDANCES.

Description of Event or Problem · 0

ADDITIONAL INFORMATION INDICATES THAT SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2023 WHEREIN THE LEAD WAS REPLACED TO ADDRESS THE ISSUE. HOWEVER, DUE TO THE PATIENT'S ANATOMY THE PHYSICIAN WAS UNABLE TO REMOVE THE NON-FUNCTIONING LEAD. LEAD WAS LEFT IN-SITU. NO FURTHER INTERVENTION IS PLANNED AT THIS TIME REGARDING THE LEAD THAT REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1848640 KIT IMPLANTABLE SLIM TIP LEAD, 50CM DRG SLIM TIP LEAD PMP ABBOTT MEDICAL MN10450-50A 6840452 05415067027153

Patients

Seq Age Sex Outcome Treatment
1 35 YR Female Other SCS IPG| SCS LEAD| SCS LEAD EXTENSION