FDA Adverse Event Injury Summary report: N

UNKNOWN

MDR report key: 17009777 · Received May 26, 2023

Report

Report Number
3002808486-2023-00162
Event Type
Injury
Date Received
May 26, 2023
Report Date
June 14, 2023
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS FOLLOW UP REPORT, WILLIAM COOK EUROPE INFORMS THAT THIS COMPLAINT HAS BEEN TRANSFERRED FROM WILLIAM COOK EUROPE TO COOK INC. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

NON-HEALTHCARE PROFESSIONAL. INVESTIGATION: THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: VENA CAVA (VC) PERFORATION, ANXIETY, FEAR, DEPRESSION, PAIN. FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. UNKNOWN IF THE REPORTED ANXIETY, FEAR, DEPRESSION, AND PAIN ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. CATALOG NUMBER AND LOT NUMBER ARE UNKNOWN. THE ALLEGED TULIP IS MANUFACTURED AND INSPECTED ACCORDING TO SPECIFICATIONS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS FOLLOW UP REPORT, WILLIAM COOK EUROPE INFORMS THAT THIS COMPLAINT HAS BEEN TRANSFERRED FROM WILLIAM COOK EUROPE TO COOK INC.

Description of Event or Problem · 0

THE FOLLOWING INFORMATION IS ALLEGED: THE PATIENT RECEIVED A GUNTHER TULIP INFERIOR VENA CAVA (IVC) FILTER ON (B)(6) 2012. APPROXIMATELY 8 YEARS, 9 MONTHS LATER A REVIEW OF THE PATIENT'S COMPUTED TOMOGRAPHY (CT) SCAN REVEALED THE HOOK OF THE COOK GUNTHER TULIP RETRIEVABLE IVC FILTER IS AT THE MID L2 VERTEBRAL BODY. THE IVC FILTER IS NOT TILTED. ALL THE STRUTS OF THE IVC FILTER PERFORATE THE IVC UP TO 12MM. ONE (1) ANTERIOR STRUT PERFORATES THE IVC WALL 12MM AND RESIDES WITHIN THE SOFT TISSUES. ONE (1) POSTERIOR STRUT PERFORATES THE IVC WALL 9MM AND CONTACTS THE L3-4 DISC. TWO (2) LATERAL STRUTS PERFORATE THE IVC WALL 9MM AND 11MM AND RESIDE WITHIN THE SOFT TISSUES. THE PATIENT EXPERIENCED ANXIETY, FEAR, DEPRESSION, AND PAIN. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED, BUT NOT YET PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1886081 UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Male Life Threatening