4D INTEGRATED TREATMENT CONSOLE
Report
- Report Number
- 2916710-2010-00055
- Event Type
- Malfunction
- Date Received
- May 21, 2010
- Date of Event
- April 7, 2010
- Report Date
- April 7, 2010
- Manufacturer
- VARIAN MEDICAL SYSTEMS
- Product Code
- IYE
- PMA / PMN Number
- K050479
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
BASED ON THE INITIAL REPORT, VARIAN'S MEDICAL ADVISOR HAS IDENTIFIED THAT THE TOTAL COURSE WAS 40 GY AND THE PLANNED DOSE FOR THAT FIELD WAS 1.3788 GY. THE MISSING WEDGE CAUSED AN ERROR OF 0.506 CGY, WHICH IS NOT CONSIDERED SIGNIFICANT AND WELL WITHIN THE TOLERANCE LIMITS. ALTHOUGH THERE WAS NO INJURY IN THIS CASE, THE AVAILABLE INFORMATION SUGGESTS A MALFUNCTION IN THE DEVICE WHICH, SHOULD IT RECUR, COULD POTENTIALLY CAUSE A SERIOUS INJURY. ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED UPON COMPLETION OF THE INVESTIGATION.
THE CUSTOMER REPORTED THAT WHEN TRYING TO CLOSE THE PATIENT AFTER TREATMENT, A SESSION SAVE ERROR MESSAGE DISPLAYED REFERRING TO A WEDGE ERROR. IT APPEARED THAT THE EDW (ENHANCED DYNAMIC WEDGE) DID NOT DELIVER AND AN OPEN FIELD WAS DELIVERED ON THE MEDIAL FIELD. THE PLAN WAS EXPORTED FROM RT CHART AFTER THE EVENT AND CONTAINED WEDGES ON BOTH FIELDS. HOWEVER, WHEN LOOKING AT THE VERSION OF THE DCM PLAN FILE EXPORTED FROM THE 4DTC (TREATMENT CONSOLE) JUST AFTER THE EVENT, THE LEFT MEDIAL FIELD WAS MISSING ITS EDW. THE ERROR WAS BROUGHT TO THE PHYSICIST'S ATTENTION WHEN THE RADIOGRAPHERS TRIED TO CLOSE THE PATIENT FOLLOWING TREATMENT AND THE TREATMENT HISTORIES WOULD NOT SAVE TO THE DATABASE. DUE TO THE MISSING EDW, THERE WAS VARIANCE IN THE PATIENT'S RADIATION PRESCRIPTION FOR THAT DAY. NO FURTHER PATIENT INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 4D INTEGRATED TREATMENT CONSOLE | LINEAR ACCELERATOR SOFTWARE | IYE | VARIAN MEDICAL SYSTEMS | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |