I-STAT CG8+ CARTRIDGE
Report
- Report Number
- 2245578-2023-00062
- Event Type
- Malfunction
- Date Received
- May 25, 2023
- Date of Event
- May 22, 2023
- Report Date
- June 14, 2023
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- CHL
- UDI-DI
- 10054749000163
- PMA / PMN Number
- K940918
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # 920523. CHEM8+: H23077, 600-9008-25, EXP:14-SEP-2023, MFG: 18-MAR-2023, GTIN: 10054749005519, 510K: K183688. CHEM8+: H23046, 600-9008-25, EXP: 14-AUG-2023, MFG: 15-FEB-2023, GTIN: 10054749005519, 510K: K183688. APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 13-JUN-2023. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AK, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOTS H23046 OR H23077 OR CG8+ CARTRIDGE LOT W22291.
ON (B)(6) 2023, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG8+ & CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT IONIZED CALCIUM RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT. THERE WERE NO TEST TIMES/DATE PRESENTED WITH THE RESULTS. METHOD: LOT#: ICA: I-STAT W22291 2.5 MMOL/L, I-STAT H23077 2.5 MMOL/L ADDITIONAL LOT USED, I-STAT H23046 2.4 MMOL/L ADDITIONAL LOT USED, VETSTAT N/A 1.21 (REF RANGE 1.25-1.5). AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 848679 | I-STAT CG8+ CARTRIDGE | CG8+ CARTRIDGE | CHL | ABBOTT POINT OF CARE | NA | W22291 | 10054749000163 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |