FDA Adverse Event Death Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 17000649 · Received May 25, 2023

Report

Report Number
1718850-2023-01004
Event Type
Death
Date Received
May 25, 2023
Report Date
May 19, 2023
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

A.1.-A.5. PATIENT INFORMATION WAS NOT PROVIDED H.10: LIVANOVA DEUTSCHLAND IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN THE UNITES STATES LIVANOVA INITIATED AN INVESTIGATION. THE PRESENT IMPORTER MEDWATCH WAS INITIALLY SUBMITTED ON 19MAY2023 AND RESUBMITTED ON 25MAY2023.

Description of Event or Problem · 0

LIVANOVA RECEIVED REPORT OF AN ALLEGED M. CHIMAERA INFECTION. ON (B)(6) 2015, PATIENT WAS OPERATED FOR ELECTIVE VALVE REPLACEMENT (OPEN HEART SURGERY) AND DISCHARGED ON (B)(6) 2015. ON (B)(6) 2022, RESULT OF PATIENT¿S SAMPLE TAKEN ON (B)(6) 2022 RESULTED POSITIVE TO MYCOBACTERIUM AVIUM COMPLEX. ON (B)(6) 202S, PATIENT ADMITTED AT (B)(6) FOR ALTERED MENTAL STATUS AND FEVER. CULTURES WERE REPEATED AND CONFIRMED INFECTION TO MYCOBACTERIUM CHIMAERA. ON (B)(6) 2022, PATIENT PASSED AWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
741380 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-82

Patients

Seq Age Sex Outcome Treatment
1 74 YR Male Death