FDA Adverse Event Malfunction Summary report: N

BIPOLAR HIGH FREQUENCY CORD

MDR report key: 16993285 · Received May 24, 2023

Report

Report Number
9610617-2023-00101
Event Type
Malfunction
Date Received
May 24, 2023
Date of Event
April 24, 2023
Report Date
May 24, 2023
Manufacturer
KARL STORZ SE & CO. KG
Product Code
GEI
UDI-DI
04048551388778
PMA / PMN Number
K171717
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

EVALUATION/INVESTIGATION OF THE PRODUCT: CONNECTOR IS DISCOLORED AND WORN FROM WEAR AND TEAR. IFU 97000149 IFU V2.0(10-2017) BIPOLAR HF CABLE WARNS "INSTRUMENTS AND ALL OF THE ACCESSORIES USED IN COMBINATION WITH THEM MUST BE CHECKED IMMEDIATELY BEFORE AND AFTER EVERY USE TO ENSURE THAT THEY ARE COMPLETE, FREE FROM DAMAGE AND TO MAKE CERTAIN THAT THE PLUG CONNECTION MAINTAINS GOOD CONTACT. POOR OR INSUFFICIENT CONTACT OR DAMAGE TO THE CABLE INSULATION AND PLUGS MAY LEAD TO VOLTAGE FLASHOVERS. THIS COMPLAINT INVOLVES A TOTAL OF 6 ITEMS (972685, 972686, 972687, 972688, 972689, 972690)OF WHICH 972687 AND 972688 ARE PROBABLY THE CAUSE OF THE REPORTED ISSUE. THOSE TWO COMPLAINTS WILL BE REPORTED TO THE LOCAL COMPETENT AUTHORITY MENTIONING THE CAUSAL COMPONENTS WITHIN THE REPORT AS INVOLVED. THERE WILL BE NO SEPARATE REPORT FOR 972685,972686, 972689,AND 972690. THE MALFUNCTION IS FILED UNDER INTERNAL COMPLAINT ID (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PRODUCT (UH801) BIPOLAR HIGH FREQUENCY CORD.ACCORDING TO THE INFORMATION RECEIVED, THE GENERATOR WAS NOT RESPONDING. A NEW RESTROSCOPE SET WAS OPENED AND THE CORD PLUGGED IN AND THE ENGERY SETTING FOR SALINE TUR WAS CHOSEN. WHEN THE SURGEON TESTED THE INSTRUMENT A LOUD POP WAS HEARD AND HE DROPPED THE INSTRUMENT INTO THE CYSTO DRAIN BAG. THE SURGEON CHANGED GLOVES AND HIS SKIN WAS FOUND TO BE INTACT. THE CORD WAS FOUND TO HAVE A BURN MARK.THERE WAS NOT HARM TO THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1508780 BIPOLAR HIGH FREQUENCY CORD BIPOLAR HIGH FREQUENCY CORD FOR KARL STORZ GENERATOR UH400U GEI KARL STORZ SE & CO. KG UH801 SO02-04342 04048551388778

Patients

Seq Age Sex Outcome Treatment
1 79 YR Male 27040EB.UH801.27040EB.UH400UUF902-T