CADSTREAM
Report
- Report Number
- 3003982333-2010-00001
- Event Type
- Other
- Date Received
- May 22, 2010
- Date of Event
- April 23, 2010
- Report Date
- May 21, 2010
- Manufacturer
- MERGE CAD INC
- Product Code
- LLZ
- PMA / PMN Number
- K081556
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE SUBURBAN IMAGING "(B)(6) FACILITY" HAD BEEN USING CADSTREAM 4.X FOR BREAST MRI REPORTING FOR APPROX 5 YEARS. THE "(B)(6) FACILITY" WAS UPGRADED TO CADSTREAM 5.0 IN (B)(6) 2009 AND THE "(B)(6) FACILITY" JUST STARTED USING CADSTREAM IN (B)(6) 2009. IN (B)(6) 2009, BOTH THE (B)(6) FACILITY CONSOLIDATED THEIR CADSTREAM SYSTEMS TO ONE CADSTREAM 5.0 ENTERPRISE SYSTEM (SERIAL (B)(4)). ALL CONFIGURATIONS FROM THE OTHER SERVERS WERE TRANSFERRED TO THIS CONSOLIDATED ENTERPRISE SERVER. THE (B)(6) FACILITY HAS BEEN VERY HAPPY WITH CADSTREAM ON THIS SERVER BUT THE (B)(6) FACILITIES REPORTED IN (B)(6) 2009, THAT THE STUDIES WERE NOT PROCESSING HOW THEY USED TO ON THEIR PRIOR CADSTREAM 4.X SYSTEM. PRODUCT SUPPORT UPDATED THE PROCESSING SETTINGS TO TRY TO MEET THE RADIOLOGIST'S REQUEST. HOWEVER, THE RADIOLOGIST CALLED AGAIN IN (B)(6) 2009 AND (B)(6) 2010, REPORTING THE PROCESSING WASN'T AS EXPECTED AND EACH TIME THE CONFIGURATION WAS MODIFIED BY SUPPORT TO TRY TO MEET THE RADIOLOGIST REQUEST. IN (B)(6) 2010, THE RADIOLOGIST AND TECH WERE TRAINED ON HOW TO SET THEIR OWN STUDY PREFERENCES AND IN THE PROCESS THEY REALIZED THAT THEY HAD THEIR PEAK SERIES SET INCORRECTLY ON THE SECOND POST WHEN IT SHOULD BE SET ON THE FIRST POST. AFTER CORRECTING THE PEAK SERIES TO BE SET TO THE FIRST POST THE PROCESSING WAS AS EXPECTED BY THE RADIOLOGIST. THE SITE DETERMINED IT NECESSARY TO RE-PROCESS AND REVIEW ALL CASES PROCESSED WITH THE IMPROPER SETTINGS. (B)(4). ALL THREE OF THESE INCIDENTS ARE BEING REPORTED TO FDA IN SEPARATE MDR REPORTS.
SOMETIME BETWEEN (B)(6) 2009 AND (B)(6) 2010, A PT WITH KNOWN MULTI-FOCAL CANCER UNDERWENT PRE-SURGICAL BREAST MRI WITH IMAGE REVIEW PERFORMED IN CADSTREAM. THE IMAGING CENTERS CADSTREAM PROCESSING SETTINGS WERE NOT SETUP TO USE THE CORRECT POST AS PEAK IN THE DYNAMIC CONTRAST ENHANCED IMAGING SERIES. THE RADIOLOGIST WAS IMPROPERLY RELYING ON THE CADSTREAM OVERLAY TO DETERMINE REGIONS OF INTEREST (OFF-LABEL USE OF DEVICE). IN (B)(6) 2010, IT WAS REALIZED THAT THE PROCESSING SETTINGS WERE NOT APPROPRIATELY SET. THE SETTINGS WERE CORRECTED AND THE STUDY RE-PROCESSED. UPON REPROCESSING IT WAS DISCOVERED THAT THE PT HAD AN ADD'L LESION. THE PT WAS ALREADY SCHEDULED FOR A MASTECTOMY, SO THIS ADD'L FINDING DID NOT IMPACT THE PT CARE OR FOLLOW-UP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CADSTREAM | LLZ | MERGE CAD INC | VERSION 5.0.7 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |