DEVICE DESCRIPTION: 4K C-MOUNT ENDOSCOPE, MITEK LOCK 30 DEGREE X 4 MM X 167 MM
Report
- Report Number
- 1221934-2023-02119
- Event Type
- Malfunction
- Date Received
- May 23, 2023
- Manufacturer
- MEDOS INTERNATIONAL SARL
- Product Code
- HRX
- UDI-DI
- 10886705032075
- PMA / PMN Number
- K080560; K98
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H10 ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS RECEIVED AT THE MANUFACTURING SITE AND EVALUATED. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: OUTER TUBE DAMAGED, DISTAL TIP. DISTAL TIP DAMAGED. DISTAL TIP FIBER DAMAGED. OPTICAL SYSTEM, OPTICAL COMPONENTS - BROKEN LENSES IN OPTICAL SYSTEM, LESS THAN 50% OF LENSES DEFECTIVE. MINOR SCRATCHES ON THE UNIT. PER SERVICE REPORTS, THIS COMPLAINT CAN BE CONFIRMED. THE DEFECTIVE PARTS NEEDS TO BE REPLACED TO RESOLVE THE ISSUES. THE FAULTY PARTS WAS IDENTIFIED AS THE ROOT CAUSE FOR THE DEVICE FAILURE DURING THE SERVICE EVALUATION. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. DEPUY MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. DEVICE HISTORY REVIEW: MANUFACTURING RECORD EVALUATION IS NOT REQUIRED AS THE REPORTED EVENT IS NOT ASSOCIATED WITH THE MANUFACTURING PROCESS AND/OR THE POTENTIAL CAUSE OF THE DEFECT CANNOT BE ASSOCIATED TO MANUFACTURING.
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H10 ADDITIONAL NARRATIVE: UDI: (B)(4). E3: REPORTER IS A J&J EMPLOYEE. H4: THE DEVICE MANUFACTURE DATE IS CURRENTLY UNAVAILABLE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED DURING SERVICE AND REPAIR EVALUATION, IT WAS DETERMINED THAT THE DEVICE WAS BROKEN/FRACTURED. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 218128 | DEVICE DESCRIPTION: 4K C-MOUNT ENDOSCOPE, MITEK LOCK 30 DEGREE X 4 MM X 167 MM | RIGID ARTHROSCOPE | HRX | MEDOS INTERNATIONAL SARL | 242268 | 10886705032075 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |