FDA Adverse Event Death Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1698672 · Received May 26, 2010

Report

Report Number
2649622-2010-05114
Event Type
Death
Date Received
May 26, 2010
Date of Event
February 25, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B) (4) OUTER INSULATION SEPARATION. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B) (4) OUTER INSULATION SEPARATION. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) OUTER INSULATION SEPARATION. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER THE PATIENT'S DEATH, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. THERE IS NO ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT THE DEATH WAS DEVICE RELATED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED. FOLLOW UP WITH THE HEALTH CARE PROFESSIONAL REVEALED THE PATIENT HAD NOT BEEN PACEMAKER DEPENDENT. THE LAST DEVICE CHECK HAD BEEN (B) (6) 2010 AND EVERYTHING WAS FINE. THE PATIENT WAS REPORTED TO HAVE BEEN IN HOSPICE.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER THE PATIENT'S DEATH, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. THERE IS NO ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT THE DEATH WAS DEVICE RELATED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER THE PATIENT'S DEATH, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. FOLLOW UP REVEALED THE PATIENT HAD NOT BEEN PACEMAKER DEPENDENT. THE LAST DEVICE CHECK HAD BEEN (B)(6) 2010 AND EVERYTHING WAS FINE. THE PATIENT WAS REPORTED TO HAVE BEEN IN HOSPICE. ADDITIONAL INFORMATION FROM THE PHYSICIAN REPORTED THE CAUSE OF DEATH WAS CHF (CONGESTIVE HEART FAILURE) AND AORTIC STENOSIS AND WAS NOT RELATED TO A MALFUNCTION OF THE GENERATOR OR LEAD SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 91 YR Death (B) (4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6947 IMPLANTABLE TACHY LEAD