OLYMPUS HIGH FREQUENCY CABLE
Report
- Report Number
- 9610773-2010-00018
- Event Type
- Death
- Date Received
- May 21, 2010
- Date of Event
- April 22, 2010
- Report Date
- April 22, 2010
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- JOS
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE RETURNED EQUIPMENTS WERE TESTED TOGETHER AND ALL OF THE DEVICES WERE FOUND TO OPERATE APPROPRIATELY. ALL OF THE DEVICES PASSED THE ELECTRICAL LEAKAGE TEST, WITH THE EXCEPTION OF THE WA00013A HF CABLE THAT FAILED THE DIELECTRIC STRENGTH TEST. THE A22002A TELESCOPE WAS FOUND TO HAVE A SLIGHT DENT ON THE PROXIMAL END AND DISTAL TIP OF THE TELESCOPE, BUT THE DENTS DID NOT AFFECT ITS FUNCTIONALITY. THE WA00013A HF CABLE WAS FORWARDED TO THE ORIGINAL EQUIPMENT MFR FOR FURTHER EVAL. BASED ON THE PRELIMINARY EVAL RESULTS, IT DOES NOT APPEAR LIKELY THAT THE DEVICES REFERENCED IN THIS REPORT CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. HOWEVER, THE CAUSE OF THE PT'S OUTCOME CANNOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFO BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL FOLLOW. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
OLYMPUS RECEIVED A REPORT THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE PT REPORTEDLY EXPERIENCED CARDIOPULMONARY ARREST. THE ANESTHESIOLOGIST AND OTHER HOSPITAL PERSONNEL ATTEMPTED TO RESUSCITATE THE PT, BUT DID NOT SUCCEED AND THE PT EXPIRED. THERE WERE NO ALLEGATIONS OF ANY DEVICE MALFUNCTION WITH THE OLYMPUS EQUIPMENT USED DURING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS HIGH FREQUENCY CABLE | HIGH FREQUENCY CABLE | JOS | OLYMPUS WINTER & IBE GMBH | WA00013A | 063 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Death | OUTER RESECTION SHEATH, MODEL A22023A,LOT 102W-001| HF RESECTION ELECTRODE, MODEL WA22557C, LOT 211209| LIGHT-GUIDE CABLE, MODEL WA03210A| EQUIPMENT INCLUDED:| SAID TO HAVE BEEN IN USE DURING THE PROCEDURE WERE| ALL OF THE FOLLOWING OLYMPUS DEVICES WHICH WERE| INNER RESECTION SHEATH,MODEL A22040A,LOT 086W-0026| ELECTROSURGICAL UNIT, MODEL UES-40, (B) (4)| UES-40 ON BIPOLAR MODE,SETTINGS: CUT-95 & COAG-140| TELESCOPE, MODEL A22002A, LOT 403030| HF RESECTION ELECTRODE, MODEL WA22306D, LOT 212261| WORKING ELEMENT, MODEL WA22367A, LOT 08ZW-0001| RETURNED TO OLYMPUS FOR EVAL. THE OLYMPUS |