FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT¿ II SYRINGE WITH DETACHED BD MICROLANCE¿ NEEDLE

MDR report key: 16981579 · Received May 22, 2023

Report

Report Number
3002682307-2023-00131
Event Type
Malfunction
Date Received
May 22, 2023
Date of Event
April 29, 2023
Report Date
August 10, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E.1 INITIAL REPORTER ADDR 1: (B)(6). H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY AS SAMPLES WERE UNAVAILABLE FOR THIS RETURN, TWENTY (20) RETAINED SAMPLES WERE OBTAINED FOR EVALUATION BY OUR INVESTIGATIVE TEAM. THE RETAINED SAMPLES WERE EXAMINED; HOWEVER, NO SIGNS OF DEFECT WERE IDENTIFIED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 300330 AND LOT NUMBER 2210509. THE REVIEW DID REVEAL ONE POTENTIALLY RELATED QUALITY NOTIFICATION DURING THE PRODUCTION PROCESS RELATED TO DAMAGED PLUNGER RODS WHICH WAS RESOLVED AT THE TIME OF DETECTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 OF THE BD DISCARDIT¿ II SYRINGES WITH DETACHED BD MICROLANCE¿ NEEDLES EXPERIENCED LEAKAGE PAST THE STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DRUG/MEDICATION LEAKAGE FROM THE BACKSIDE OF PLUNGER WHILE ASPIRATING DRUGS.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 OF THE BD DISCARDIT¿ II SYRINGES WITH DETACHED BD MICROLANCE¿ NEEDLES EXPERIENCED LEAKAGE PAST THE STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DRUG/MEDICATION LEAKAGE FROM THE BACKSIDE OF PLUNGER WHILE ASPIRATING DRUGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182434 BD DISCARDIT¿ II SYRINGE WITH DETACHED BD MICROLANCE¿ NEEDLE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2210509

Patients

Seq Age Sex Outcome Treatment
1 Unknown