BD PLASTIPAK ¿ - 3-PIECE SYRINGE
Report
- Report Number
- 3003152976-2023-00193
- Event Type
- Malfunction
- Date Received
- May 22, 2023
- Date of Event
- April 27, 2023
- Report Date
- June 12, 2023
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY: PHOTOS RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL EVALUATION, IT CAN BE OBSERVED TWO STRIPS JOINED BY THE SHORT EDGE, AND EMPTY PACKAGE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2202063, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. BASED ON THE TEAMS INVESTIGATION, POSSIBLE ROOT CAUSE IS ASSOCIATED WITH MALADJUSTMENT IN THE CROSSCUT OF BLISTERS, IF STRIPS OF UNITARY PACKAGING ARE NOT CORRECTLY SEPARATED, AN EMPTY STRIP IS NOT CORRECTLY DISCARDED AND THEREFORE, IT CAN BE PACKAGED IN THE CASE.
IT WAS REPORTED THAT 2 BD PLASTIPAK ¿ - 3-PIECE SYRINGE EXPERIENCED POOR PERFORATION ON PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE IS NO BREAKAGE GROOVE IN THE PACKAGED SYRINGES AND THE SECOND PACKAGING PART IS ALWAYS EMPTY(SECOND SYRINGE MISSING). MEANS THERE ARE MUCH LESS SYRINGES IN THE TOTAL PACKAGE THAN THE TOTAL PACKAGE DESCRIBES.
IT WAS REPORTED THAT 2 BD PLASTIPAK ¿ - 3-PIECE SYRINGE EXPERIENCED POOR PERFORATION ON PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE IS NO BREAKAGE GROOVE IN THE PACKAGED SYRINGES AND THE SECOND PACKAGING PART IS ALWAYS EMPTY(SECOND SYRINGE MISSING). MEANS THERE ARE MUCH LESS SYRINGES IN THE TOTAL PACKAGE THAN THE TOTAL PACKAGE DESCRIBES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 849084 | BD PLASTIPAK ¿ - 3-PIECE SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2202063 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |