FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK ¿ - 3-PIECE SYRINGE

MDR report key: 16981269 · Received May 22, 2023

Report

Report Number
3003152976-2023-00193
Event Type
Malfunction
Date Received
May 22, 2023
Date of Event
April 27, 2023
Report Date
June 12, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: PHOTOS RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL EVALUATION, IT CAN BE OBSERVED TWO STRIPS JOINED BY THE SHORT EDGE, AND EMPTY PACKAGE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2202063, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. BASED ON THE TEAMS INVESTIGATION, POSSIBLE ROOT CAUSE IS ASSOCIATED WITH MALADJUSTMENT IN THE CROSSCUT OF BLISTERS, IF STRIPS OF UNITARY PACKAGING ARE NOT CORRECTLY SEPARATED, AN EMPTY STRIP IS NOT CORRECTLY DISCARDED AND THEREFORE, IT CAN BE PACKAGED IN THE CASE.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 BD PLASTIPAK ¿ - 3-PIECE SYRINGE EXPERIENCED POOR PERFORATION ON PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE IS NO BREAKAGE GROOVE IN THE PACKAGED SYRINGES AND THE SECOND PACKAGING PART IS ALWAYS EMPTY(SECOND SYRINGE MISSING). MEANS THERE ARE MUCH LESS SYRINGES IN THE TOTAL PACKAGE THAN THE TOTAL PACKAGE DESCRIBES.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 BD PLASTIPAK ¿ - 3-PIECE SYRINGE EXPERIENCED POOR PERFORATION ON PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE IS NO BREAKAGE GROOVE IN THE PACKAGED SYRINGES AND THE SECOND PACKAGING PART IS ALWAYS EMPTY(SECOND SYRINGE MISSING). MEANS THERE ARE MUCH LESS SYRINGES IN THE TOTAL PACKAGE THAN THE TOTAL PACKAGE DESCRIBES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
849084 BD PLASTIPAK ¿ - 3-PIECE SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2202063

Patients

Seq Age Sex Outcome Treatment
1 Unknown