FDA Adverse Event Injury Summary report: N

IMP TM 3.7MM MTX FULL,10M

MDR report key: 16974990 · Received May 22, 2023

Report

Report Number
0002023141-2023-01394
Event Type
Injury
Date Received
May 22, 2023
Date of Event
May 8, 2023
Report Date
November 15, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024018990
PMA / PMN Number
K132258
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER(B)(4). A1: PATIENT IDENTIFIER UNKNOWN / NOT PROVIDED. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. ZIMVIE RECEIVED THE REPORTED IMPLANT FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED, A FRACTURE HAS BEEN IDENTIFIED ON THE IMPLANTS BODY. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1241937. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1241937 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WERE MISSING OR CONFUSING INSTRUCTIONS FOR USE AND CLINICIAN SELECTS IMPLANT THAT IS UNABLE TO RESIST LONG-TERM OCCLUSAL LOADING. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR. THE FRACTURE HAS BEEN IDENTIFIED ON THE BODY. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #29 LOST INTEGRATION AND WAS REMOVED DUE TO A FRACTURE. THE PATIENT CAME IN FOR A TORQUE TEST AND THE IMPLANT WAS BROKEN AND BURIED THE WHOLE TIME.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
760731 IMP TM 3.7MM MTX FULL,10M DENTAL IMPLANT DZE ZIMMER DENTAL TMTB10 1241937 00889024018990

Patients

Seq Age Sex Outcome Treatment
1 58 YR Prefer Not To Disclose Required Intervention