STAT DL 9.5 FR. 40 CC. IAB
Report
- Report Number
- 2248146-1998-00557
- Event Type
- Malfunction
- Date Received
- May 22, 1998
- Date of Event
- September 29, 1997
- Report Date
- May 15, 1998
- Manufacturer
- DATASCOPE CORP.
- Product Code
- DSP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
(THIS FOLLOW-UP MDR WAS MAILED TO THE FDA ON 07/08/1998). EVALUATION: UNDERWATER LEAK TESTING DISCLOSED A LEAK IN THE IAB MEMBRANE. UNDER MICROSCOPY, A SMOOTH & SQUARE EDGED PENETRATION WAS SEEN AT THE LEAK SITE. PROBABLE CAUSE OF DIFFICULTY: THE PHYSICAL EVIDENCE INDICATES THAT THE SOURCE OF THE BALLOON LEAK WAS A PENETRATION OF THE MEMBRANE, PROBABLY CAUSED BY CONTACT WITH A SHARP EDGED OBJECT. THE MEMBRANE DAMAGE MAY HAVE OCCURRED PRIOR TO OR DURING BALLOON INSERTION.
THE IAB LEAKED DURING INSERTION. ON 06/10/1998, THE FOLLOWING WAS REPORTED TO DATASCOPE: THE DOCTOR PERFORMED CORONARY ARTERY BYPASS GRAFT SURGERY ON A 70 YEAR OLD PT WITH 100% LEFT MAIN LESION. PT HAD A HEMODYNAMIC COMPROMISE AFTER BIENG WEANED OFF CARDIOPULMONARY BYPASS, AND IT WAS DECIDED TO SUPPORT HIM ON IABP. THE DOCTOR INTRODUCED THE SHEATH BY SELDINGER TECHNIQUE INTO THE RIGHT FEMORAL ARTERY. WHEN THE BALLOON WAS INTRODUCED INTO THE SHEATH OVER THE GUIDE WIRE FOR JUST 2-4 INCHES, THE DOCTOR COULD SEE BLOOD INSIDE THE BALLOON. BLOOD WAS REGURGITATING INTO THE BALLOON. THE IAB WAS REMOVED AND ANOTHER WAS INSERTED. THERE WAS NO PT INJURY OR COMPLICATION AS A RESULT OF THE EVENT. (MULTIPLE COMPLAINT). (EVENT COMPLICATIONS): UNKNOWN - REPORTED 05/15/1998; NONE - REPORTED 06/10/1998. (PT'S CURRENT STATUS): UNK -05/15/1998, 06/10/1998.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STAT DL 9.5 FR. 40 CC. IAB | INTRA-AORTIC BALLOON CATHETER | DSP | DATASCOPE CORP. | 0684-00-0283 | 03/25/99 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |