FDA Adverse Event Injury Summary report: N

WHISTLE TIP URETERAL CATHETER

MDR report key: 1696663 · Received May 21, 2010

Report

Report Number
1825146-2010-00017
Event Type
Injury
Date Received
May 21, 2010
Date of Event
April 22, 2010
Report Date
April 22, 2010
Manufacturer
COOK UROLOGICAL, INC.
Product Code
EYB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

ONE USED CATHETER WAS RECEIVED IN TWO PIECES IN A SPECIMEN JAR. THE DISTAL SEGMENT MEASURES 17.4CM AND THE PROXIMAL SEGMENT MEASURES 54.9CM AND THE TWO PIECES ARE MATING FRACTURES NOTING THE ENTIRE CATHETER WAS RETURNED. THE SEPARATION HAS THE APPEARANCE OF BEING CAUGHT ON AN INSTRUMENT OF UNDETERMINED ORIGIN CUTTING THE CATHETER.

Description of Event or Problem · 1

UPON REMOVING THE CATHETER FOLLOWING THE PROCEDURE, THE CATHETER BROKE OFF INSIDE THE PT. THE SEGMENT WAS REMOVED FROM THE URETER DURING THE SAME PROCEDURE. INITIAL INFO INDICATES SURGICAL PROCEDURE TO REMOVE. ADDITIONAL INFO CONCERNING THE INCIDENT AS WELL AS REMOVAL HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WHISTLE TIP URETERAL CATHETER EYB CATHETER, URETERAL, GASTRO-UROLOGY EYB COOK UROLOGICAL, INC. NA U1766282

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention