WHISTLE TIP URETERAL CATHETER
MDR report key: 1696663
·
Received May 21, 2010
Report
- Report Number
- 1825146-2010-00017
- Event Type
- Injury
- Date Received
- May 21, 2010
- Date of Event
- April 22, 2010
- Report Date
- April 22, 2010
- Manufacturer
- COOK UROLOGICAL, INC.
- Product Code
- EYB
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
ONE USED CATHETER WAS RECEIVED IN TWO PIECES IN A SPECIMEN JAR. THE DISTAL SEGMENT MEASURES 17.4CM AND THE PROXIMAL SEGMENT MEASURES 54.9CM AND THE TWO PIECES ARE MATING FRACTURES NOTING THE ENTIRE CATHETER WAS RETURNED. THE SEPARATION HAS THE APPEARANCE OF BEING CAUGHT ON AN INSTRUMENT OF UNDETERMINED ORIGIN CUTTING THE CATHETER.
Description of Event or Problem · 1
UPON REMOVING THE CATHETER FOLLOWING THE PROCEDURE, THE CATHETER BROKE OFF INSIDE THE PT. THE SEGMENT WAS REMOVED FROM THE URETER DURING THE SAME PROCEDURE. INITIAL INFO INDICATES SURGICAL PROCEDURE TO REMOVE. ADDITIONAL INFO CONCERNING THE INCIDENT AS WELL AS REMOVAL HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WHISTLE TIP URETERAL CATHETER | EYB CATHETER, URETERAL, GASTRO-UROLOGY | EYB | COOK UROLOGICAL, INC. | NA | U1766282 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |