FDA Adverse Event
Malfunction
Summary report: N
AU400 CHEMISTRY SYSTEM
MDR report key: 1696504
·
Received May 24, 2010
Report
- Report Number
- 2050012-2010-00293
- Event Type
- Malfunction
- Date Received
- May 24, 2010
- Date of Event
- May 3, 2010
- Report Date
- May 24, 2010
- Manufacturer
- BECKMAN COULTER MISHIMA K.K.
- Product Code
- JJE
- PMA / PMN Number
- K981473
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE GAS SPRINGS FOR THE TOP LID WERE CHECKED ON SEMI-ANNUAL PREVENTIVE MAINTENANCE (PM) IN (B)(4) 2009. AT THE TIME, THE GAS SPRING DID NOT FAIL AS PER THE PM PROCEDURE. A FIELD SERVICE ENGINEER (FSE) WAS ON-SITE FOR THIS EVENT AND REPLACED THE GAS SPRING.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING AN INCIDENT WITH THE USE OF THE AU400 CHEMISTRY SYSTEM. THE CUSTOMER STATED THAT THE LID FELL ON HER HEAD DURING A STARTUP. THERE WAS NO INJURY AND NO MEDICAL ATTENTION REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AU400 CHEMISTRY SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER MISHIMA K.K. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |