T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2023-70795
- Event Type
- Malfunction
- Date Received
- May 19, 2023
- Date of Event
- April 16, 2023
- Report Date
- July 13, 2023
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613762
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED AND EVALUATION IS PERFORMED. PER TANDEM USER GUIDE: ONLY USE U-100 HUMALOG OR U-100 NOVOLOG WITH YOUR PUMP. ONLY U-100 HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE PUMP. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE HYPOGLYCEMIA (LOW BG) OR HYPERGLYCEMIA (HIGH BG) EVENTS. PER TANDEM USER GUIDE: AVOID EXPOSURE OF YOUR PUMP TO TEMPERATURES BELOW 40°F (5°C) OR ABOVE 99°F (37°C). INSULIN CAN FREEZE AT LOW TEMPERATURES OR DEGRADE AT HIGH TEMPERATURES. INSULIN THAT HAS BEEN EXPOSED TO CONDITIONS OUTSIDE OF THE MANUFACTURER¿S RECOMMENDED RANGES CAN AFFECT THE SAFETY AND PERFORMANCE OF THE PUMP. H3 OTHER TEXT : DEVICE NOT RETURNED.
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. REPORTEDLY THE CUSTOMER IS USING APIDRA INSULIN., AND USING COLD INSULIN. TANDEM TECHNICAL SUPPORT (CTS) INFORMED CUSTOMER THAT USING APIDRA INSULIN, AND COLD INSULIN, IS OFF LABEL PER THE USER GUIDE. CUSTOMER CHANGED THE PUMP SUPPLIES AND SUCCESSFULLY RESUMED INSULIN THERAPY, AND CUSTOMER HAS MANUAL INSULIN INJECTIONS IF NEEDED. CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS 250-460 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1017304 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | W0740114 | 00850006613762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Female | INFUSION SET: AUTOSOFT 30INSULIN TYPE: APIDRA |