FDA Adverse Event Injury Summary report: N

PHILIPS MOBILEDIAGNOST WDR 2.2

MDR report key: 16961011 · Received May 18, 2023

Report

Report Number
MW5117662
Event Type
Injury
Date Received
May 18, 2023
Date of Event
May 15, 2023
Report Date
May 16, 2023
Manufacturer
PHILIPS MEDICAL SYSTEMS INTERNATIONAL BV/SEDECAL S.A.
Product Code
IZL
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Description of Event or Problem · 0

ON (B)(6) 2023, ALL PHILIPS DIGITAL DIAGNOST PORTABLE X-RAYS ACROSS THE COUNTRY CEASED FUNCTIONING WITHOUT WARNING. ACROSS THE MCLAREN SYSTEM, 12 UNITS AT SIX DIFFERENT HOSPITALS WERE AFFECTED. THE LARGEST IMPACT WAS AT (B)(6) WHERE ALL UNITS FAILED AND PATIENT CARE WAS SIGNIFICANTLY AFFECTED AS NO PORTABLE X-RAYS WERE ABLE TO BE CONDUCTED FOR 2 DAYS. ON 5/16, PHILIPS INDICATED THE CAUSE OF THE ISSUES WAS IDENTIFIED AS A CERTIFICATE PROTECTING THE ELEVA USER LOGIN WHICH EXPIRED. THE PROPOSED SOLUTION IS TO UPGRADE TO SOFTWARE 2.2.4, BUT NO SCHEDULE HAS BEEN PROVIDED TO REMEDIATE OUR SYSTEMS. (B)(6) REQUESTED THE CAPA INFORMATION ON THIS ISSUE, AND THE CORRECTIVE ACTION WAS CLEAR, HOWEVER (B)(6) WAS UNABLE TO GET A PREVENTIVE ACTION PLAN TO ENSURE THIS EVENT DOESN'T OCCUR AGAIN. (B)(6) IS REQUESTING FDA OPEN AN INVESTIGATION INTO THIS NATIONAL ISSUE AND ENSURE PHILIPS HAS THE QUALITY MANAGEMENT SYSTEM AND APPROPRIATE CUSTOMER COMMUNICATIONS IN PLACE TO AVOID THIS TYPE OF ISSUE IN THE FUTURE. PLANS SHOULD ENSURE IMPLEMENTATION OF SAFEGUARDS SO THAT ENTIRE PRODUCT LINES ACROSS THE COUNTRY AREN'T RENDERED USELESS AT ANY TIME WITHOUT WARNING. THE PATIENT IMPACT OF THIS OUTAGE WILL CONTINUE UNTIL THE SOFTWARE UPDATE IS COMPLETE. REFERENCE REPORTS: MW5117657, MW5117658, MW5117659, MW5117660, MW5117661, MW5117663, MW5117664, MW5117665, MW5117666, MW5117667, MW5117668.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1409952 PHILIPS MOBILEDIAGNOST WDR 2.2 SYSTEM, X-RAY, MOBILE IZL PHILIPS MEDICAL SYSTEMS INTERNATIONAL BV/SEDECAL S.A.

Patients

Seq Age Sex Outcome Treatment
1 Unknown Life Threatening