FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 16959457 · Received May 18, 2023

Report

Report Number
3004753838-2023-101397
Event Type
Malfunction
Date Received
May 18, 2023
Date of Event
April 29, 2023
Report Date
June 30, 2023
Manufacturer
DEXCOM, INC.
Product Code
QBJ
UDI-DI
00386270001627
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Additional Manufacturer Narrative · 0

CAPA-000367. B5: DESCRIBE EVENT OR PROBLEM - CORRECTION G3: DATE RECEIVED BY MANUFACTURER - ADDITIONAL H2 TYPE OF FOLLOW UP - CORRECTION H6: INVESTIGATION CONCLUSION - CORRECTION.

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND PASSED. NORDIC BLUETOOTH PAIRING TEST WAS PERFORMED AND PASSED. A REVIEW OF THE BIN FILE WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643614 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 9445-24 5300542 00386270001627

Patients

Seq Age Sex Outcome Treatment
1 20 YR Male