FDA Adverse Event Malfunction Summary report: N

BD BBL¿ MUELLER HINTON AGAR WITH 5% SHEEP BLOOD

MDR report key: 16959436 · Received May 18, 2023

Report

Report Number
1119779-2023-00577
Event Type
Malfunction
Date Received
May 18, 2023
Date of Event
May 1, 2023
Report Date
September 5, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JTZ
UDI-DI
10382902212758
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: DURING MANUFACTURING OF MATERIAL 221176, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD FOR BATCH 3073513 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. ALL BATCHES ARE TESTED PRIOR TO RELEASE AND RESULTS REPORTED ON THE CERTIFICATE OF ANALYSIS WHICH CAN BE OBTAINED AT WWW.BD.COM/REGDOCS. MUELLER HINTON II AGAR IS STABILITY TESTED BIENNIALLY FOR BIOLOGICAL PERFORMANCE TO ENSURE SATISFACTORY PERFORMANCE THROUGHOUT SHELF LIFE WITH THE ORGANISMS THAT ARE REPORTED ON THE CERTIFICATE OF ANALYSIS. ALL PERFORMANCE TESTING ON THIS BATCH WAS SATISFACTORY AT THE TIME OF RELEASE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 3073513. RETENTION SAMPLES FROM BATCH 3073513 WERE NOT AVAILABLE FOR INSPECTION. TWO RETURN SHIPMENTS OF BATCH 3073513 SAMPLES WERE RECEIVED FOR INVESTIGATION. ONE UNOPENED SLEEVE (TEN PLATES) WAS RECEIVED ON MAY 15, 2023 IN A BIOHAZARD SHIPPING KIT (TIME STAMP 1509). ANOTHER UNOPENED SLEEVE (10 PLATES) WAS RECEIVED ON MAY 22, 2023 IN AN INSULATED BOX WITH ICE PACKS AND RETURNS FOR THREE OTHER BATCHES (TIME STAMP 1510). RETURN SAMPLES WERE PH TESTED AND THE AVERAGE PH OF SIX PLATES WAS 7.3. THE PH SPECIFICATION FOR THIS PRODUCT IS 7.2 TO 7.4 WHEN TAKEN AT 23 TO 27 DEGREES C. NO PHOTOS WERE RECEIVED FOR INVESTIGATION. THE PH OF THE PRODUCT MAY VARY DEPENDING ON THE AGE OF THE PRODUCT AND THE TYPE OF PH METER AND PROBE USED, HOWEVER RETURN SAMPLES TESTED WITHIN PH SPECIFICATIONS. THIS COMPLAINT CANNOT BE CONFIRMED. BD WILL CONTINUE TO TREND COMPLAINTS FOR PH VARIANCE.

Additional Manufacturer Narrative · 0

D2A: COMMON DEVICE NAME: CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ MUELLER HINTON AGAR WITH 5% SHEEP BLOOD HAS HAD 3 LOW PH READS ON DIFFERENT OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORT MEDIA 221176 LOW PH WITH THREE READING TAKEN 7.06 ,7.11,7.13.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ MUELLER HINTON AGAR WITH 5% SHEEP BLOOD HAS HAD 3 LOW PH READS ON DIFFERENT OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORT MEDIA 221176 LOW PH WITH THREE READING TAKEN 7.06 ,7.11,7.13.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
478731 BD BBL¿ MUELLER HINTON AGAR WITH 5% SHEEP BLOOD SEE H.10 JTZ BECTON, DICKINSON & CO. (SPARKS) 221275 3073513 10382902212758

Patients

Seq Age Sex Outcome Treatment
1 Unknown