BASE PLATE UNCEMENTED 20 MM POST LENGTH
Report
- Report Number
- 0001822565-2023-01188
- Event Type
- Injury
- Date Received
- May 18, 2023
- Date of Event
- December 23, 2022
- Report Date
- May 11, 2023
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- PHX
- UDI-DI
- 00889024489448
- PMA / PMN Number
- K172767
- Removal / Correction Number
- NI
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). THIS WAS INITIALLY REPORTED ON 0001822565-2023-00148; HOWEVER, PRODUCT ID NEEDED TO BE UPDATED BASED ON ADDITIONAL INFORMATION RECEIVED. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. RADIOGRAPHS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE AVAILABLE RECORDS IDENTIFIED THE FOLLOWING: RELATIVE RADIOLUCENCY OF THE GLENOID SUGGESTS SEVERE OSTEOPENIA. HOWEVER, NO DEFINITE FRACTURE CAN BE SEEN. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. ASSOCIATED PRODUCTS AND REPORTS: 0001822565-2023-00148-1; ITEM#00434901213; LOT#65538166. 0001822565-2023-00149-1; ITEM#00-4360-036-00; LOT#65356130. 0001822565-2023-00152; ITEM#00434903603; LOT#65352850. OTHER ASSOCIATED PRODUCTS: ITEM#47436108300; LOT#56672638. ITEM#0104223033; LOT#3021582. ITEM#0104223042; LOT#3127010. ITEM#47430906125; LOT#65267726. ITEM#47430904601; LOT#65157507. ITEM#47430902501; LOT#65553821.
IT WAS REPORTED PATIENT UNDERWENT A RIGHT SHOULDER REVISION APPROXIMATELY TWO (2) WEEKS POST IMPLANTATION DUE POOR BONE QUALITY, DISLOCATION, AND FIXATION OF IMPLANT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 643592 | BASE PLATE UNCEMENTED 20 MM POST LENGTH | SHOULDER PROSTHESIS, REVERSE CONFIGURATION | PHX | ZIMMER BIOMET, INC. | NI | 64390359 | 00889024489448 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Required Intervention| H |