FDA Adverse Event Other Summary report: N

COMPACT WS125C

MDR report key: 1695917 · Received May 17, 2010

Report

Report Number
1718850-2010-00101
Event Type
Other
Date Received
May 17, 2010
Date of Event
April 27, 2010
Report Date
April 27, 2010
Manufacturer
SORIN GROUP ITALIA
Product Code
CAC
PMA / PMN Number
K982650
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SORIN GROUP (B)(4) MANUFACTURES THE COMPACT WASH SET. THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF SORIN GROUP (B)(4). THE TECHNICIAN REPORTED THAT AT THE END OF PROCESSING AUTOLOGOUS BLOOD AND DURING THE EMPTY MODE, AN ERROR MESSAGE DISPLAYED ON THE COMPACT MACHINE. AS THE TECHNICIAN POWER CYCLED THE MACHINE, A POPPING SOUND WAS HEARD. THE TUBING LINE ON THE OUTLET OF THE RESERVOIR SEPARATED, SPRAYING BLOOD IN THE ROOM, ON THE TECHNICIAN AND IN THE STERILE FIELD. THE PROCESSING SET WAS INADVERTENTLY DISCARDED BY THE HOSPITAL. THE CAUSE OF THE INCIDENT CAN NOT BE DETERMINED WITHOUT THE PHYSICAL PRODUCT FOR EVALUATION. NO FURTHER ACTION IS DEEMED NECESSARY AT THIS TIME. THERE WAS PERSONNEL BLOOD EXPOSURE DURING THIS INCIDENT. THIS MEDWATCH REPORT IS BEING FILED AS A RESULT OF THE EXPOSURE.

Description of Event or Problem · 1

THE TECHNICIAN REPORTED THAT AT THE END OF PROCESSING AUTOLOGOUS BLOOD AND DURING THE EMPTY MODE, AN ERROR MESSAGE DISPLAYED ON THE COMPACT MACHINE. AS THE TECHNICIAN POWER CYCLED THE MACHINE, A POPPING SOUND WAS HEARD. THE TUBING LINE ON THE OUTLET OF THE RESERVOIR SEPARATED, SPRAYING BLOOD IN THE ROOM, ON THE TECHNICIAN AND IN THE STERILE FIELD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPACT WS125C APPARATUS, AUTOTRANSFUSION CAC SORIN GROUP ITALIA NA 1001200103

Patients

Seq Age Sex Outcome Treatment
1