VITROS CHEMISTRY PRODUCTS BENZ REAGENT
Report
- Report Number
- 1319808-2010-00011
- Event Type
- Malfunction
- Date Received
- May 21, 2010
- Date of Event
- March 12, 2010
- Report Date
- May 21, 2010
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JXM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION INTO THIS EVENT CONCLUDED THAT A FALSE NEGATIVE BENZ RESULT OCCURRED. THE MOST LIKELY CAUSE IS A COMBINATION OF A KNOWN LIMITATION OF THE VITROS BENZ REAGENT RELATED TO LOW CROSS REACTIVITY WITH LORAZEPAM AND USE ERROR IN THE INTERPRETATION OF A VITROS BENZ RESULT. THE PROFICIENCY SAMPLE CONTAINED LORAZEPAM AND AT TARGET CONCENTRATION OF 1000 NG/ML AND THE VITROS BENZ REAGENT IDENTIFIED A BENZODIAZEPINE COMPOUND AT A MUCH LOWER CONCENTRATION OF 260 NG/ML. THE INTENDED USE SECTION OF THE VITROS BENZ IFU STATES THAT THE VITROS CHEMISTRY PRODUCT BENZ ASSAY IS INTENDED FOR USE BY PROFESSIONAL LABORATORY PERSONNEL. IT PROVIDES ONLY A PRELIMINARY TEST RESULT. A MORE SPECIFIC ALTERNATIVE CHEMICAL METHOD MUST BE USED TO CONFIRM A RESULT OBTAINED WITH THE VITROS BENZ ASSAY. GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGMENT SHOULD BE APPLIED TO ANY DRUG-OF-ABUSE TEST RESULT. NO MALFUNCTION OF THE VITROS BENZ REAGENT OCCURRED. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT.
A FALSE NEGATIVE VITROS BENZ RESULT WAS OBTAINED FROM A PROFICIENCY SAMPLE WHILE USING THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE MAGNITUDE AND DIRECTION OF THE FALSE NEGATIVE VITROS BENZ RESULT OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF IT OCCURRED ON A PATIENT SAMPLE. PATIENT SAMPLES WERE NOT AFFECTED AND THERE WAS NO ALLEGATION OF PATIENT HARM. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS CHEMISTRY PRODUCTS BENZ REAGENT | IN-VITRO DIAGNOSTIC | JXM | ORTHO-CLINICAL DIAGNOSTICS | 1523-09-9381 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |