FDA Adverse Event Malfunction Summary report: N

VITROS CHEMISTRY PRODUCTS BENZ REAGENT

MDR report key: 1695522 · Received May 21, 2010

Report

Report Number
1319808-2010-00011
Event Type
Malfunction
Date Received
May 21, 2010
Date of Event
March 12, 2010
Report Date
May 21, 2010
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
JXM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION INTO THIS EVENT CONCLUDED THAT A FALSE NEGATIVE BENZ RESULT OCCURRED. THE MOST LIKELY CAUSE IS A COMBINATION OF A KNOWN LIMITATION OF THE VITROS BENZ REAGENT RELATED TO LOW CROSS REACTIVITY WITH LORAZEPAM AND USE ERROR IN THE INTERPRETATION OF A VITROS BENZ RESULT. THE PROFICIENCY SAMPLE CONTAINED LORAZEPAM AND AT TARGET CONCENTRATION OF 1000 NG/ML AND THE VITROS BENZ REAGENT IDENTIFIED A BENZODIAZEPINE COMPOUND AT A MUCH LOWER CONCENTRATION OF 260 NG/ML. THE INTENDED USE SECTION OF THE VITROS BENZ IFU STATES THAT THE VITROS CHEMISTRY PRODUCT BENZ ASSAY IS INTENDED FOR USE BY PROFESSIONAL LABORATORY PERSONNEL. IT PROVIDES ONLY A PRELIMINARY TEST RESULT. A MORE SPECIFIC ALTERNATIVE CHEMICAL METHOD MUST BE USED TO CONFIRM A RESULT OBTAINED WITH THE VITROS BENZ ASSAY. GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGMENT SHOULD BE APPLIED TO ANY DRUG-OF-ABUSE TEST RESULT. NO MALFUNCTION OF THE VITROS BENZ REAGENT OCCURRED. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

A FALSE NEGATIVE VITROS BENZ RESULT WAS OBTAINED FROM A PROFICIENCY SAMPLE WHILE USING THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE MAGNITUDE AND DIRECTION OF THE FALSE NEGATIVE VITROS BENZ RESULT OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF IT OCCURRED ON A PATIENT SAMPLE. PATIENT SAMPLES WERE NOT AFFECTED AND THERE WAS NO ALLEGATION OF PATIENT HARM. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS CHEMISTRY PRODUCTS BENZ REAGENT IN-VITRO DIAGNOSTIC JXM ORTHO-CLINICAL DIAGNOSTICS 1523-09-9381

Patients

Seq Age Sex Outcome Treatment
1