FDA Adverse Event Death Summary report: N

CAPSUREFIX

MDR report key: 1695397 · Received May 21, 2010

Report

Report Number
2649622-2010-05106
Event Type
Death
Date Received
May 21, 2010
Date of Event
July 25, 2008
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS INFORMATION WAS RECEIVED FROM A COMPETITOR. ALL DATA PERTINENT TO THE EVENT IS PROVIDED. EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION WAS BREACHED AND HAD A DEPRESSION. IT WAS NOTED THAT THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION.

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS INFORMATION WAS RECEIVED FROM A COMPETITOR. ALL DATA PERTINENT TO THE EVENT IS PROVIDED. EVALUATION SUMMARY (B) (4) OUTER INSULATION SEPARATION. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER FOLLOWING THE PATIENT'S DEATH, WAS ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. THERE IS NO ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT THE DEATH WAS DEVICE RELATED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED. FOLLOW UP WITH THE CLINIC REPORTED THE PATIENT WAS LAST SEEN IN 2008 WHEN THE DEVICE WAS CHECKED.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER FOLLOWING THE PATIENT'S DEATH, WAS ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. THERE IS NO ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT THE DEATH WAS DEVICE RELATED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5068 ASKU

Patients

Seq Age Sex Outcome Treatment
1 96 YR Death| O