FDA Adverse Event
Injury
Summary report: N
PITT EASY ABUTMENT
MDR report key: 1695289
·
Received May 20, 2010
Report
- Report Number
- 9611820-2010-00008
- Event Type
- Injury
- Date Received
- May 20, 2010
- Report Date
- April 22, 2010
- Manufacturer
- SYBRON IMPLANT SOLUTIONS GMBH
- Product Code
- NHA
- PMA / PMN Number
- K053242
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
ON (B) (4), 2010, THE DOCTOR REPORTED THAT AN ABUTMENT HEX SCREW FRACTURED. THE EVALUATION OF THE RETURNED PRODUCT INDICATED THAT THE HEX SCREW FRACTURED AS A RESULT OF OVERLOADING. THE ABUTMENT WAS USED AS A POST FOR A SINGLE TOOTH RESTORATION IN THE REGION OF THE FIRST MOLAR, WHICH RESULTED IN OVERSTRESSING THE HEX SCREW AND THE SUBSEQUENT FRACTURE. THE PRODUCT MET SPECIFICATIONS. THIS IS NOT A PRODUCT FAILURE.
Description of Event or Problem · 1
ON (B) (6), 2010 A DOCTOR REPORTED TO (B) (4) THAT A PITT EASY ABUTMENT HEX SCREW FRACTURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PITT EASY ABUTMENT | ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS | NHA | SYBRON IMPLANT SOLUTIONS GMBH | 823506 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other| R |