FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1695289 · Received May 20, 2010

Report

Report Number
9611820-2010-00008
Event Type
Injury
Date Received
May 20, 2010
Report Date
April 22, 2010
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

ON (B) (4), 2010, THE DOCTOR REPORTED THAT AN ABUTMENT HEX SCREW FRACTURED. THE EVALUATION OF THE RETURNED PRODUCT INDICATED THAT THE HEX SCREW FRACTURED AS A RESULT OF OVERLOADING. THE ABUTMENT WAS USED AS A POST FOR A SINGLE TOOTH RESTORATION IN THE REGION OF THE FIRST MOLAR, WHICH RESULTED IN OVERSTRESSING THE HEX SCREW AND THE SUBSEQUENT FRACTURE. THE PRODUCT MET SPECIFICATIONS. THIS IS NOT A PRODUCT FAILURE.

Description of Event or Problem · 1

ON (B) (6), 2010 A DOCTOR REPORTED TO (B) (4) THAT A PITT EASY ABUTMENT HEX SCREW FRACTURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH 823506

Patients

Seq Age Sex Outcome Treatment
1 Other| R