FDA Adverse Event Malfunction Summary report: N

PRECISE DESKTOP PRO

MDR report key: 1694899 · Received May 14, 2010

Report

Report Number
9617016-2010-00002
Event Type
Malfunction
Date Received
May 14, 2010
Date of Event
April 26, 2010
Report Date
May 14, 2010
Manufacturer
ELEKTA LTD.
Product Code
IYE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION INTO THIS SITUATION IS ONGOING AT THIS TIME. ONCE THE INVESTIGATION IS COMPLETED, MORE DETAILS AND A CONCLUSION WILL BE PROVIDED.

Description of Event or Problem · 1

WHEN A PT IS LOADED FROM MOSAIQ SOFTWARE AND TREATED NORMALLY WITH XXX MU, AFTER TREATMENT IS FINISHED, THE DETAILS ARE SENT TO MOSAIQ USING ICOM. DURING THE TRANSMISSION OF DETAILS FROM DESKTOP PRO TO MOSAIQ, IF ANOTHER PT IS SELECTED ON THE XVI SYS, THIS XVI INFO IS PRIORITIZED AND THE DATA FROM THE END OF THE TREATMENT IS NEVER REC'D BY MOSAIQ. IF THIS HAPPENS, AS MOSAIQ HAS NOT BEEN ABLE TO RECORD THE TREATMENT, IT PRESENTS THE CLINICAL USER WITH THE FOLLOWING WARNING DIALOGUE: "MU DELIVERED IS LESS THAN EXPECTED, XXX MU REMAINING, DELIVER REMAINDER?" IF THE USER SELECTS 'YES' THEN MOSAIQ WILL DOWNLOAD THE REMAINING MU, (WHICH IT CALCULATES INCORRECTLY AS THE TOTAL FIELD MU OF THE PREVIOUSLY DELIVERED FIELD), ALLOWING AN IMMEDIATE RETREAT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISE DESKTOP PRO DESKTOP PRO IYE ELEKTA LTD. ALL

Patients

Seq Age Sex Outcome Treatment
1