ELEKTA MEDICAL LINEAR ACCELERATOR
Report
- Report Number
- 3015232217-2023-00030
- Event Type
- Malfunction
- Date Received
- May 17, 2023
- Date of Event
- April 17, 2023
- Report Date
- October 25, 2023
- Manufacturer
- ELEKTA SOLUTIONS AB
- Product Code
- IYE
- UDI-DI
- 05060191071574
- PMA / PMN Number
- K210500
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE MANUFACTURER'S INVESTIGATION IS ON-GOING AND FURTHER INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS COMPLETED.
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
THE CUSTOMER REPORTED A RATTLING SOUND WHEN THE GANTRY ROTATES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 501194 | ELEKTA MEDICAL LINEAR ACCELERATOR | ACCELERATOR, LINEAR, MEDICAL | IYE | ELEKTA SOLUTIONS AB | 156780 | 05060191071574 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |