ENTERALITE INFINITY ENTREAL FEEDING PUMP
Report
- Report Number
- 1722139-2010-00060
- Event Type
- Other
- Date Received
- May 6, 2010
- Date of Event
- April 9, 2010
- Report Date
- May 5, 2010
- Manufacturer
- MOOG MEDICAL DEVICES GROUP
- Product Code
- LZH
- PMA / PMN Number
- K031199
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
METHOD: ACTUAL DEVICE WAS EVALUATED. RESULTS: THE EVALUATION INCLUDED PERFORMANCE TESTING (ACCURACY, VERIFICATION OF AIR/NO FOOD ALARM, ETC). MULTIPLE FORMULAS AND SETTINGS WERE USED TO RECREATE THE FAILURE EXPERIENCED BY THE CUSTOMER. THE PUMP PERFORMED ACCORDING TO SPECIFICATION DURING EACH RUN (ALARMED AND SHUT OFF AFTER FOOD WAS GONE). CONCLUSION: THE ALLEGED COMPLAINT/DEFECT COULD NOT BE DUPLICATED. THE PUMP PERFORMED ACCORDING TO SPECIFICATION. WE ARE ATTEMPTING TO CONTACT THE CUSTOMER/PATIENT TO OBTAIN MORE INFORMATION. IF MORE CUSTOMER/PATIENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT MAY BE SUBMITTED.
REPORTED BY THE CUSTOMER AS: PUMP CONTINUED TO PUMP AFTER FOOD WAS GONE. CUSTOMER STATED "NO "DOSE DONE" OR "NO FOOD" ALARM. NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENTERALITE INFINITY ENTREAL FEEDING PUMP | LZH | MOOG MEDICAL DEVICES GROUP | INFINITY PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |