KIT DRAIN 100CC W/7MM PERF
Report
- Report Number
- 1423537-2023-00937
- Event Type
- Malfunction
- Date Received
- May 16, 2023
- Date of Event
- April 19, 2023
- Report Date
- May 16, 2023
- Manufacturer
- PR006 PUERTO RICO-ANASCO
- Product Code
- GCY
- UDI-DI
- 10192253022537
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT INDICATES THAT THE LOT NUMBER IS UNKNOWN, ALTHOUGH THE CUSTOMER HAS TWO LOT NUMBERS 1200283 AND 1200569 ON THE SHELF. THEREFORE, THE DEVICE HISTORY RECORD WAS REVIEWED FOR THE TWO LOT NUMBERS. THESE LOT NUMBERS SHOWED THE PRODUCT WAS INSPECTED AND RELEASED IN COMPLIANCE WITH ALL REQUIREMENTS. THE SAMPLE WAS NOT RETURNED FOR INVESTIGATION; THEREFORE, AN EVALUATION OF THE COMPLAINT DEVICE FOR DEFICIENCY OF CONSTRUCTION COULD NOT BE PERFORMED AND THE ROOT CAUSE REMAINS UNKNOWN. ALTHOUGH NO SAMPLE WAS RECEIVED, INSTRUCTIONS FOR USE EVALUATION WAS PERFORMED TO CONFIRM THAT THE REQUIRED STEPS TO AVOID BREAKAGE OF THE DRAIN ARE AVAILABLE.
CUSTOMER REPORTED THAT THE SPINE SURGEON TRIED TO PLACE A JP DRAIN AND WHEN HE GRABBED THE END TO PULL IT THROUGH THE INCISION, THE PLASTIC LITERALLY CRUMBLED. PIECE WAS RETRIEVED WITH NO ADVERSE EFFECT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1380986 | KIT DRAIN 100CC W/7MM PERF | APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED | GCY | PR006 PUERTO RICO-ANASCO | SU130-1360 | UNKNOWN | 10192253022537 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Female |