FDA Adverse Event Malfunction Summary report: N

KIT DRAIN 100CC W/7MM PERF

MDR report key: 16942850 · Received May 16, 2023

Report

Report Number
1423537-2023-00937
Event Type
Malfunction
Date Received
May 16, 2023
Date of Event
April 19, 2023
Report Date
May 16, 2023
Manufacturer
PR006 PUERTO RICO-ANASCO
Product Code
GCY
UDI-DI
10192253022537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT INDICATES THAT THE LOT NUMBER IS UNKNOWN, ALTHOUGH THE CUSTOMER HAS TWO LOT NUMBERS 1200283 AND 1200569 ON THE SHELF. THEREFORE, THE DEVICE HISTORY RECORD WAS REVIEWED FOR THE TWO LOT NUMBERS. THESE LOT NUMBERS SHOWED THE PRODUCT WAS INSPECTED AND RELEASED IN COMPLIANCE WITH ALL REQUIREMENTS. THE SAMPLE WAS NOT RETURNED FOR INVESTIGATION; THEREFORE, AN EVALUATION OF THE COMPLAINT DEVICE FOR DEFICIENCY OF CONSTRUCTION COULD NOT BE PERFORMED AND THE ROOT CAUSE REMAINS UNKNOWN. ALTHOUGH NO SAMPLE WAS RECEIVED, INSTRUCTIONS FOR USE EVALUATION WAS PERFORMED TO CONFIRM THAT THE REQUIRED STEPS TO AVOID BREAKAGE OF THE DRAIN ARE AVAILABLE.

Description of Event or Problem · 0

CUSTOMER REPORTED THAT THE SPINE SURGEON TRIED TO PLACE A JP DRAIN AND WHEN HE GRABBED THE END TO PULL IT THROUGH THE INCISION, THE PLASTIC LITERALLY CRUMBLED. PIECE WAS RETRIEVED WITH NO ADVERSE EFFECT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1380986 KIT DRAIN 100CC W/7MM PERF APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED GCY PR006 PUERTO RICO-ANASCO SU130-1360 UNKNOWN 10192253022537

Patients

Seq Age Sex Outcome Treatment
1 83 YR Female